Setting Up the Oscilloscope
Technical Specifications
1-8
Technical Specifications
Below are the technical specifications for the Infiniium 9000A Series oscilloscopes.
Vertical: Analog Channels
Analog Bandwidth (-3 db) 50
Ω* 9064A: 600 MHz
9104A: 1.0 GHz,
9254A: 2.5 GHz,
9404A: 4.0 GHz
Calculated Rise Time / Fall Time,
10%-90% at 50
Ω
9064A; 540 ps
9104A: 253 ps,
9254A: 142 ps,
9404A: 85 ps
Calculated Rise Time / Fall Time,
20%-80% at 50
Ω
9064A: 310 ps
9104A: 174 ps,
9254A: 98 ps,
9404A: 59 ps
Input channels DSO models: 4 analog
MSO models: 4 analog + 16 digital
Input impedance* 50
Ω \2.5%, 1 MΩ \1% (11 pF typical)
Input sensitivity 1
ΜΩ: 1 mV/div to 5 V/div
50 Ω: 1 mV/ to 1 V/div
Input coupling 1
ΜΩ: AC (3.5 Hz), DC
50 Ω: DC
Hardware bandwidth limit 20 MHz on 1 ΜΩ input only
Vertical resolution** 8 bits, >=12 bits with averaging
Channel-to-channel isolation DC to 50 MHz: 50 dB
>50 MHz to 1 GHz: 40 dB
>1 GHz to 4 GHz: 25 dB
DC gain accuracy**
\2% of full scale at full resolution on channel scale \5
o
C from cal temp
*Denotes warranted specification, all others are typical. Specifications are valid only after a 30 minute warm-up period
and
\5
o
C from firmware calibration temperature.
**Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale