
Setting Up the Oscilloscope
To verify basic operation for the 9000A series oscilloscope
1-24
The display will pause momentarily while the oscilloscope adjusts the sweep speed
and vertical scale. You should then see a square wave with peak-to-peak amplitude
of approximately 5 divisions and about four cycles on screen. If you do not see the
waveform, ensure your power source is adequate, the oscilloscope is properly
powered-on, and the probe is connected securely to the front-panel channel input BNC
and to the probe calibration output.
5 Move the mouse around the mouse surface and verify that the on screen pointer
follows the mouse movement.
6 Touch the pointer of the touch screen stylus to the surface of the screen and
move it around while verifying that the pointer follows the movement. Ensure
that the front panel Touchscreen button is illuminated (symbolizing that the
touch screen is enabled). You do not need to use the stylus in many instances
when controlling the oscilloscope via the touch screen - a finger is effective
as well.
Commenti su questo manuale