
Setting Up the Oscilloscope
To verify basic operation for the 9000A series oscilloscope
1-23
To verify basic operation for the 9000A series oscilloscope
1 Connect one end of the calibration cable using SMA to precision BNC
adaptors to channel 1.
2 Connect the other end of the calibration cable to the front panel connector with
the square wave label.
Figure 1-7
Verifying Basic Oscilloscope Operation for 9000A Series Oscilloscopes.
3 Press the Default Setup key on the front panel.
The display will pause momentarily while the oscilloscope is configured to its default
settings.
4 Press the Autoscale key on the front panel.
Front panel
connector
with square
wave label
Commenti su questo manuale