Setting Up the Oscilloscope
Technical Specifications
1-9
Maximum input voltage* 1 ΜΩ: 150 Vrms or dc, CAT I, \250 V (dc + ac) in ac coupling
50 Ω: 5 Vrms
Offset range 1 M
Ω
(vertical sensitivity: offset)
1 mV to < 10 mV/div: \2 V
10 mV to < 20 mV/div: \5 V
20 mV to <100 mV/div: \10 V
100 mV to 1 V/div: \20 V
1 V to 5 V/div: \100 V
50 Ω
\
12 div or \4 V, whichever is smallest
Offset accuracy**
\(1.25% of channel offset + 1% of full scale + 1 mV)
Dynamic range 1 ΜΩ: \8 div from center screen
50 Ω: \8 div from center screen
DC voltage measurement
accuracy*, **, ***
Dual Cursor:
\[(dc gain accuracy) + resolution]
Single Cursor: \[(dc gain accuracy) + offset accuracy + (resolution/2)]
Vertical: Digital Channels (on all MSO models)
Input channels 16 digital channels
Threshold groupings Pod 1: D7 - D0
Pod 2: D15 - D8
User-defined threshold range
\8.00 V in 10 mV increments
Maximum input voltage
\40 V peak, CAT I
Threshold accuracy
\(100 mV + 3% of threshold setting)
Input dynamic range
\10 V about threshold
Minimum input voltage swing 500 mV peak-to-peak
Input impedance (flying leads) 100 K
Ω \2% (app. 8 pF) at probe tip
Resolution 1 bit
Vertical: Analog Channels
*Denotes warranted specification, all others are typical. Specifications are valid only after a 30 minute warm-up period
and
\5
o
C from firmware calibration temperature.
**Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale
***Calculated from the bandwidth