Command Reference 5
Programmer’s Guide 95
Marker Subsection
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
Turns off all markers on all the traces.
Key access: Marker
> All Off
Continuous Peaking Marker
:CALCulate:MARKer[1]|2|3|4|5|6:CPEak[:STATe]
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:CPEak[:STATe]?
Toggles the continuous peak search function between on and off.
Example: :CALCulate:MARKer1:CPEak ON
Key access: Shift > Peak > More > Continuous PK
Frequency Counter Marker
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt[:STATe]
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt[:STATe]?
Turns on/off the marker frequency counter. To query the frequency
counter, use :CALCulate:MARKer[1]:FCOunt:X? If the
specified marker number is not the active marker, it becomes the active
marker. If the specified marker number is not on, it is turned on and
becomes the active marker. A 1 is returned only if marker count is on
and the selected number is the active marker.
*RST: Off
Remarks: If a frequency count x value is generated when the frequency count state
is off, then 0 is returned.
Example: :CALCulate:MARKer1:FCOunt 1
Key access: Marker > Function > Counter