Agilent Technologies N9344C Manuale di Servizio Pagina 213

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 224
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 212
Command Reference 5
Programmer’s Guide 205
Set Power Meter Video Bandwidth Type
[:SENSe]:BANDwidth|BWIDth HIGH|MEDium|LOW|OFF
[:SENSe]:BANDwidth|BWIDth?
Use this command to toggle the video bandwidth type between high,
middle, low and off. It’s only available for U2020 X series peak power
sensor.
Example: :SENSe:BANDwidth HIGH
Key access: Meas > General Setup > More 1 of 2 > Video B/W
Set Power Meter Video Average Count
[:SENSe]:AVERage:VIDeo:COUNt <value>
[:SENSe]:AVERage:VIDeo:COUNt?
This command defines the average count number of power meter. It’s
only available for U2020X series peak power sensor, and the trigger
type must be set to continuos trigger or single trigger. The count number
parameter can only be set to N-th power of 2.
[:SENSe]:AVERage:VIDeo:COUNt:AUTO OFF|ON|0|1
[:SENSe]:AVERage:VIDeo:COUNt:AUTO?
Use this command to turn on/off the average count of power meter.
Example: :SENSe:AVERage:VIDeo:COUNt 16
Key access: Meas > General Setup > More 1 of 2 > Video Avg
Set Trace X Start Time
:SENSe:TRACe:OFFSet:TIME <value>
:SENSe:TRACe:OFFSet:TIME?
This command defines the trace X start time of peak power sensor.
Range: 0 to 1 s
Example: :SENSe:TRACe:OFFSet:TIME 0.15
Key access: Meas > Peak Setup > Trace Setup > X Start