172 Programmer’s Guide
5 Command Reference
AM Depth Upper Limit
:CALCulate:AM:LIMit:MDEPth:UPPer <value>
:CALCulate:AM:LIMit:MDEPth:UPPer?
Use this command to set the upper limit of AM depth.
Range: 0.1% to 100%
Example: :CALCulate:AM:LIMit:MDEPth:UPPer 50
Key access: Shift > Limit> AMod Depth Up
AM Depth Lower Limit
:CALCulate:AM:LIMit:MDEPth:LOWer <value>
:CALCulate:AM:LIMit:MDEPth:LOWer?
Use this command to set the lower limit of AM depth.
Range: 0.1% to 100%
Example: :CALCulate:AM:LIMit:MDEPth:LOWer 10
Key access: Shift > Limit > AMod Depth Low
AM Carrier Frequency Offset Upper Limit
:CALCulate:AMA:LIMit:FOFFset:UPPer <freq>
:CALCulate:AMA:LIMit:FOFFset:UPPer?
Use this command to turn on/off peak hold state of AM detector.
Range: –1MHz to +1MHz
Example: :CALCulate:AMA:LIMit:FOFFset:UPPer 1kHz
Key access: Shift > Limit > CarrFreqOfStUp