Command Reference 5
Programmer’s Guide 141
CAT Trace Math Type
[:SENSe]:CAT:TRACe:MATH OFF|ADD|SUB|DIV
[:SENSe]:CAT:TRACe:MATH?
This command toggles the trace math type of cable and antenna
measurement. CAT option support three trace type, measurement trace
data plus memory saved trace data, immurement trace data minus
memory saved trace data, and memory saved data minus measurement
trace data.
Example: :CAT:TRACe:MATH ADD
Key access: Trace > Trace Math
CAT Trace Average Number
[:SENSe]:CAT:AVERage:COUNt <value>
[:SENSe]:CAT:AVERage:COUNt?
This command defines the trace average number of CAT test.
[:SENSe]:CAT:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CAT:AVERage[:STATe]?
Use this command to turn on/off the trace average operation.
Example: :CAT:AVERage:COUNt 20
Key access: Trace > Average Number
CAT Trace Data Format
:FORMat[:TRACe][:DATA] ASCii|REAL
:FORMat[:TRACe][:DATA]?
This command toggles the trace data format. This command affects only
return trace data format. ASC stands for ASCII string be separated by
comma, and REAL stands for 4 bytes length float without comma.
Range: ASCii, REAL
Example: :FORMat ASC