Agilent Technologies N9344C Manuale di Servizio Pagina 132

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 224
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 131
124 Programmer’s Guide
5 Command Reference
SEM Offset Fail Mask
[:SENSe]:SEMask:OFFSet:LIST:TEST ABSolute|RELa-
tive|AND|OR,ABSolute|RELative|AND|OR,ABSo-
lute|RELative|AND|OR,ABSolute|RELative|AND|OR,AB
Solute|RELative|AND|OR
[:SENSe]:SEMask:OFFSet:LIST:TEST?
Use this command to choose the mask type in SEM measurement.
Range: ABSolute, RELative, AND, OR
Example: :SENSe:SEMask:OFFSet:LIST:TEST
ABSolute,ABSolute, RELative, AND, OR
Key access: Meas > SEM > Offset/Limit > More 1 of 3 > More 2 of 3 > Fail Mask
SEM Meas Type
[:SENSe]:SEMask:TYPE TPRef|PSDRef
[:SENSe]:SEMask:TYPE?
Use this command to switch the measurement type in SEM
measurement.
Example: SENSe:SEMask:TYPE PSDRef
Key access: Meas > SEM > Meas Type
SEM Lower Offset Fail Flag
[:SENSe]:SEMask:OFFSet:LIST:LOWer:FAIL?
Use this command to query fail flag of each lower offset channel in
SEM measurement.
Example: :SEMask:OFFSet:LIST:LOWer:FAIL?
SEM Lower Offset Power
[:SENSe]:SEMask:OFFSet:LIST:LOWer:POWer?
Use this command to query the measured power of each lower offset
channel in SEM measurement.
Example: :SEMask:OFFSet:LIST:LOWer:POWer?