Command Reference 5
Programmer’s Guide 155
Modify Channel By Custom For List Sweep
:CHSCanner:EDIT:LIST:MODify:CUSTom
<index>,<freq>,<band-
width>{,<index>,<freq>,<bandwidth>}
This command edit the sweep channel by custom, and it requires to
input the channel index, new center frequency and new bandwidth.
Example: :CHSCanner:EDIT:LIST:MODify:CUSTom
1,10000000,10000,2,20000000,20000
Key access: Meas > List Edit > Edit > Edit Method > Custom
Modify Channel By Signal Standard for List Sweep
:CHSCanner:EDIT:LIST:MODify:STD <index>,<signal
std.>,<ch. no>{,<index>,<signal std.>,<ch. no>}
This command modify scan channels according to the build-in signal
standard. The address of build-in signal stand file is “d:/”.
Example: :CHSCanner:EDIT:LIST:MODify:STD
“1,d:/CDMA-AWS_DL.STD,13”
Key access: Meas > List Edit > > Edit > Edit Method Std> Signal Std
Delete Channel For List Sweep
:CHSCanner:EDIT:LIST:DELete <index>{,<index>}
This command deletes channels for list sweep mode.
Example: :CHSCanner:EDIT:LIST:DELete 1
Key access: Meas > List Edit > Delete
Delete All Channels For List Sweep
:CHSCanner:EDIT:LIST:DELete:ALL
This command delete all channels for list sweep mode.
Example: :CHSCanner:EDIT:LIST:DELete:ALL
Key access: Meas > List Edit > Delete All