U2701A/U2702A
Performance characteristics
[1]
(continued)
Vertical system: oscilloscope channels (continued)
Single-cursor accuracy ±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~½ LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.1(10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)}
= ±6.385 mV
Dual-cursor accuracy ±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.4% (80 mV)}
= ±3.52 mV
Horizontal
Range 1 ns/div to 50 s/div
Time base accuracy 20 ppm
Delay range
Pre-trigger: –100 %
Post-trigger: +100 %
Modes Main, roll, XY
XY Yes
Reference position Center
Trigger System
Sources Ch 1, Ch 2, Ext (not applicable for TV trigger)
Modes Normal, single, auto trigger
Holdoff time 60 ns
Selections Edge, pulse width, TV
Edge Triggers on a rising or falling edge, alternating, or either edge of any source
Pulse width Triggers on a pulse width greater than, equal to, or less than a specified time limit, with
time limits ranging from 16 ns to 10 s.
Minimum lower limit: 8 ns
Minimum upper limit: 16 ns
Maximum pulse width setting: 10 s
TV Triggers on one of three standard television waveforms: NTSC, PAL, SECAM
TV trigger sensitivity: 0.6 division of sync signal. Modes supported include Field 1, Field 2,
all fields, or any line within a field.
Autoscale Single-button automatic setup of all channels
Oscilloscope channel triggering
Range (internal) ±4 div from center screen
Coupling AC (< 15 Hz)
LF reject (~ 35 kHz)
HF reject (~ 35 kHz)
External (EXT) triggering
Input impedance 1 MW: ≈ 16 pF
Maximum input CAT I 30 Vrms, 42 Vpk
Range DC coupling: trigger level ±1.25 V and ±2.5 V
EXT trigger pulse width > 2.5 ns
[1]
All specifications are warranted. specifications are valid after a 30-minute warm-up and within ±100 ºC of last calibration temperature.