61
U2741A
[1]
Frequency measurement can only be done in auto range mode. Specifications are for 30 minutes warm-up time, using one second aperture. Measuring
method is using reciprocal counting technique with AC coupled input at AC voltage function. Gate time of 0.1 second or 1 second.
[2]
Only applicable for square wave measurement.
Frequency specifications
[1]
Frequency accuracy
Function Range
Accuracy
(% of reading + % of
range)
Minimum input
frequency
Temperature
coefficient (% of range)
Frequency 20 Hz to 300 kHz 0.0200 + 0.003 1 Hz 0.005
Frequency Sensitivity for AC Voltage
Function Range
Minimum sensitivity (RMS sine wave)
Frequency (Hz)
20 ~ 100k 100k ~ 300k
AC voltage
100 mV
[2]
20 mV 20 mV
1 V 100 mV 120 mV
10 V 1 V 1.2 V
100 V 10 V 20 V
250 V 100 V 120 V
Temperature specifications
Temperature accuracy
Function Thermistor type Range Accuracy Temperature coefficient
Temperature 5 kΩ thermistor
–80.0 °C to 150 °C
Probe accuracy + 0.2 °C 0.002 °C
–112 °F to 302 °F
NOTE To ensure better measurement results and to guard against the change of environment or setup, always
enable the Null offset.