U2701A/U2702A
Performance characteristics
[1]
(continued)
External (EXT) triggering (continued)
Trigger level sensitivity For ±1.25 V range setting:
DC to 100 MHz: 100 μV
> 100 MHz: 200 μV
For ±2.5 V range setting:
DC to 100 MHz: 250 μV
> 100 MHz: 500 μV
Display
Interpolation Sin(x)/x
Display types Dots and vectors
Persistence Off, infinite
Format XY, roll
Measurement features
Automatic measurements Measurements are continuously updated.
Cursors track last selected measurement.
Voltage Peak-to-peak, maximum, minimum, average, amplitude, top, base, Vrms,
overshoot, preshoot, crest, standard deviation, cycle RMS, RMS AC
Time Frequency, period, +width, –width, +duty cycle, –duty cycle, rise time, fall time, delay,
phase
Frequency Maximum peak
Cursors Modes: Manual
Type: Time, voltage and frequency (FFT)
Measurements: DT, DV, frequency, Peak Scan (FFT), DPeak
Math functions Add, substract, multiply, FFT, divide
FFT
Points 1250 points (for 500 ns and above)
Source of FFT Source channels 1 or 2
Window Hanning, Hamming, Blackman-Harris, Rectangular, Flattop
Noise floor –50 dB to –90 dB depending on averaging
Amplitude Display in dBV
Maximum frequency 250 MHz
[1]
All specifications are warranted. specifications are valid after a 30-minute warm-up and within ±100 ºC of last calibration temperature.