56
U2722A/U2723A
Performance characteristics
Rise/fall time (ms)
[1]
For resistive measurement
[2]
±1 μA 170.0 15.0
±10 μA 18.0 5.0
±100 μA 6.0 1.0
±1 mA 1.0 1.0
±10 mA 1.0 1.0
±120 mA 1.0 1.0
Remote sense operating range Ensure that the maximum voltage between the OUTPUT+ and
SENSE+, OUTPUT–, and SENSE– does not exceed 3 V.
Temperature coeffi cient Maximum change in output/readback per °C after a 30-minute
warm-up is 0.15.
Guard output resistance 0.2 kΩ
Noise 10 Hz to 20 MHz (peak-peak) 100 mV typical into a resistive load (fl oating mode).
Output voltage overshoot, ±(% of output + offset)
[2]
During turn-on or turn-off, the output plus overshoot < 0.1% + 10 mV.
Programming language SCPI (Standard Commands for Programmable Instruments)
Maximum sense lead resistance 1 MΩ for rated accuracy
Voltage line regulation 0.01% of range
Voltage load regulation 0.01% + 100 μV
Current line regulation 0.04% of range
Current load regulation 0.04% + 100 μA
Recommended calibration interval One year
[1]
Drive 50% of 1 V or 10 V output with a resistive load. Rise time is from 10% to 90% of program voltage change at maximum current. Fall time is
from 90% to 10% of program voltage change at maximum current.
[2]
Measurements obtained are per default bandwidth setting.
U2722A U2723A
Voltage range Current range
1 μA 10 μA 100 μA 1 mA 10 mA 120 mA
2 V 50 mV 50 mV 50 mV 50 mV 30 mV 30 mV
20 V 50 mV 50 mV 50 mV 50 mV 30 mV 30 mV
Noise 10 Hz to 20 MHz (Peak-peak)
NOTE All channels are isolated from the ground and from each other. Isolation is +60 VDC, Category 1.•
All specifications are based on three hours warm-up time. •
The measurement accuracy value is x ( 1 + a * y ), where, •
x = accuracy specification at room temperature, ◦
a = temperature coefficient, and ◦
y = temperature change from room temperature in °C ◦