U2722A/U2723A
Product specifications
General
Model U2722A/U2723A
Number of outputs 3
Output ratings (at 0 °C to 50 °C)
Voltage –20 V to 20 V
Current –120 mA to 120 mA
Performance specification
U2722A/U2723A Range Accuracy
[1]
Resolution
Voltage programming
12 months (at 25 °C ± 3 °C), ±(% of output + offset)
±2 V 0.075% + 1.5 mV 0.1 mV
±20 V 0.05% + 10 mV 1 mV
Current programming
12 months (at 25 °C ± 3 °C), ±(% of output + offset)
±1 μA 0.085% + 0.85 nA 100 pA
±10 μA 0.085% + 8.5 nA 1 nA
±100 μA 0.075% + 75 nA 10 nA
±1 mA 0.075% + 750 nA 100 nA
±10 mA 0.075% + 7.5 μA 1 μA
±120 mA 0.1% + 100 μA 20 μA
Voltage readback
12 months (over USB with respect to the actual output at
25 °C ± 3 °C), ±(% of output + offset)
±2 V 0.075% + 1.5 mV 0.1 mV
±20 V 0.05% + 10 mV 1 mV
Current readback
12 months (over USB with respect to the actual output at
25 °C ± 3 °C), ±(% of output + offset)
±1 μA 0.085% + 0.85 nA 100 pA
±10 μA 0.085% + 8.5 nA 1 nA
±100 μA 0.075% + 75 nA 10 nA
±1 mA 0.075% + 750 nA 100 nA
±10 mA 0.075% + 7.5 μA 1 μA
±120 mA 0.1% + 100 μA 20 μA
[1]
Accuracy measurements are based on NPLC 10.