Agilent Technologies 2000 X-Series Manuale di Servizio Pagina 132

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 930
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 131
132 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
5 Common (*) Commands
*OPC (Operation Complete)
(see page 790)
Command Syntax
*OPC
The *OPC command sets the operation complete bit in the Standard Event
Status Register when all pending device operations have finished.
Query Syntax
*OPC?
The *OPC? query places an ASCII "1" in the output queue when all
pending device operations have completed. The interface hangs until this
query returns.
Return Format
<complete><NL>
<complete> ::= 1
See Also "Introduction to Common (*) Commands" on page 123
"*ESE (Standard Event Status Enable)" on page 126
"*ESR (Standard Event Status Register)" on page 128
"*CLS (Clear Status)" on page 125
Vedere la pagina 131
1 2 ... 127 128 129 130 131 132 133 134 135 136 137 ... 929 930

Commenti su questo manuale

Nessun commento