Agilent Technologies 2000 X-Series Manuale di Servizio Pagina 388

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 930
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 387
388 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
21 :MTESt Commands
:MTESt:COUNt:FWAVeforms
(see page 790)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and
all waveforms collected on the channel specified by the optional parameter
or collected on the currently specified source channel (:MTESt:SOURce) if
there is no parameter.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:COUNt:WAVeforms" on page 391
":MTESt:COUNt:TIME" on page 390
":MTESt:COUNt:RESet" on page 389
":MTESt:SOURce" on page 409
Example Code "Example Code" on page 379
Vedere la pagina 387
1 2 ... 383 384 385 386 387 388 389 390 391 392 393 ... 929 930

Commenti su questo manuale

Nessun commento