504 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh
(see page 790)
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh <width>
<width> ::= integer from 4 to 64 in NR1 format
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh command sets the width
of the SPI data pattern anywhere from 4 bits to 64 bits.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh?
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh? query returns the
current SPI data pattern width setting.
Return Format
<width><NL>
<width> ::= integer from 4 to 64 in NR1 format
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA" on page 503
":SBUS<n>:SPI:SOURce:MISO" on page 501
NOTE
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh should be set before
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA.