474 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:IIC:TRIGger:PATTern:DATA
(see page 790)
Command Syntax
:SBUS<n>:IIC:TRIGger:PATTern:DATA <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :SBUS<n>:IIC:TRIGger:PATTern:DATA command sets IIC data. The data
value can range from 0x00 to 0x0FF (hexadecimal). Use the don't care
data pattern (- 1 or 0xFFFFFFFF) to ignore the data value.
Query Syntax
:SBUS<n>:IIC:TRIGger:PATTern:DATA?
The :SBUS<n>:IIC:TRIGger:PATTern:DATA? query returns the current
pattern for IIC data.
Return Format
<value><NL>
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:IIC:TRIGger:PATTern:ADDRess" on page 473
":SBUS<n>:IIC:TRIGger:PATTern:DATa2" on page 475
":SBUS<n>:IIC:TRIGger[:TYPE]" on page 477