Agilent Technologies 2000 X-Series Manuale di Servizio Pagina 528

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 930
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 527
528 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:QUALifier
(see page 790)
Command Syntax
:SBUS<n>:UART:TRIGger:QUALifier <value>
<value> ::= {EQUal | NOTequal | GREaterthan | LESSthan}
The :SBUS<n>:UART:TRIGger:QUALifier command selects the data qualifier
when :TYPE is set to RDATa, RD1, RD0, RDX, TDATa, TD1, TD0, or TDX
for the trigger when in UART mode.
Query Syntax
:SBUS<n>:UART:TRIGger:QUALifier?
The :SBUS<n>:UART:TRIGger:QUALifier? query returns the current UART
trigger qualifier.
Return Format
<value><NL>
<value> ::= {EQU | NOT | GRE | LESS}
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:UART:TRIGger:TYPE" on page 529
Vedere la pagina 527
1 2 ... 523 524 525 526 527 528 529 530 531 532 533 ... 929 930

Commenti su questo manuale

Nessun commento