Agilent Technologies 2000 X-Series Manuale di Servizio Pagina 344

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 930
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 343
344 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:DEFine
(see page 790)
Command Syntax
:MEASure:DEFine <meas_spec>
<meas_spec> ::= {DELay | THResholds}
The :MEASure:DEFine command sets up the definition for measurements
by specifying the delta time or threshold values. Changing these values
may affect the results of other measure commands. The table below
identifies which measurement results that can be affected by redefining
the DELay specification or the THResholds values. For example, changing
the THResholds definition from the default 10%, 50%, and 90% values may
change the returned measurement result.
:MEASure:DEFine
DELay Command
Syntax
:MEASure:DEFine DELay,<delay spec>
<delay spec> ::= <edge_spec1>,<edge_spec2>
<edge_spec1> ::= [<slope>]<occurrence>
<edge_spec2> ::= [<slope>]<occurrence>
<slope> ::= {+ | -}
<occurrence> ::= integer
MEASure Command DELay THResholds
DUTYcycle x
DELay x x
FALLtime x
FREQuency x
NWIDth x
OVERshoot x
PERiod x
PHASe x
PREShoot x
PWIDth x
RISetime x
VAVerage x
VRMS x
Vedere la pagina 343
1 2 ... 339 340 341 342 343 344 345 346 347 348 349 ... 929 930

Commenti su questo manuale

Nessun commento