:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 495
:SBUS<n>:SPI:BITorder
(see page 790)
Command Syntax
:SBUS<n>:SPI:BITorder <order>
<order> ::= {LSBFirst | MSBFirst}
The :SBUS<n>:SPI:BITorder command selects the bit order, most significant
bit first (MSB) or least significant bit first (LSB), used when displaying
data in the serial decode waveform and in the Lister.
Query Syntax
:SBUS<n>:SPI:BITorder?
The :SBUS<n>:SPI:BITorder? query returns the current SPI decode bit
order.
Return Format
<order><NL>
<order> ::= {LSBF | MSBF}
Errors "- 241, Hardware missing" on page 753