E8362/3/4C
Test port output
Description Specification Supplemental information
Standard 014 UNL UNL and 014
Frequency range
E8362C 10 MHz to 20 GHz
E8363C 10 MHz to 40 GHz
E8364C 10 MHz to 50 GHz
Nominal power
2
E8362C 0 dBm -5 dBm -5 dBm -5 dBm
E8363/4C -12 dBm -17 dBm -17 dBm -17 dBm
Frequency resolution 1 Hz 1 Hz 1 Hz 1 Hz
CW accuracy ± 1ppm ± 1ppm ± 1ppm ± 1ppm
Frequency stability ±1 ppm, 0 to 40 °C, typical
±0.2 ppm/yr, typical
Power level accuracy
1
10 to 45 MHz
3
±2.0 dB ±2.0 dB ±2.0 dB ±2.0 dB
45 MHz to 10 GHz ±1.5 dB ±1.5 dB ±1.5 dB ±1.5 dB Variation from nominal
10 to 20 GHz ±2.0 dB ±2.0 dB ±2.0 dB ±2.0 dB power in range 0
20 to 40 GHz ±3.0 dB ±3.0 dB ±3.0 dB ±3.0 dB (step attenuator at 0 dB).
40 to 45 GHz ±3.0 dB ±3.5 dB ±3.0 dB ±3.5 dB
45 to 50 GHz ±3.0 dB ±4.0 dB ±3.0 dB ±4.0 dB
Power level linearity
6
10 to 45 MHz
3
±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
45 MHz to 20 GHz ±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
Test reference is at the
20 to 40 GHz ±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
nominal power level
40 to 50 GHz ±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
±1.0 dB
4
(step attenuator at 0 dB).
Power range
1, 5, 7
10 to 45 MHz
3
-25 to +2 dB -25 to +2 dBm -87 to +2 dBm -87 to +2 dBm
45 MHz to 10 GHz -25 to +5 dB -25 to +5 dBm -87 to +3 dBm -87 to +3 dBm
8
10 to 20 GHz -24 to +3 dB -25 to +2 dBm -86 to +1 dBm -87 to 0 dBm
9
20 to 30 GHz -23 to 0 dBm -25 to -2 dBm -85 to -2 dBm -87 to -4 dBm
30 to 40 GHz -23 to -4 dBm -25 to - 6 dBm -85 to -6 dBm -87 to -8 dBm
40 to 45 GHz -25 to -5 dBm -27 to -7 dBm -87 to -9 dBm -87 to -11 dBm
45 to 50 GHz -25 to -10 dBm -27 to -12 dBm -87 to -15 dBm -87 to -17 dBm
Power sweep range (ALC)
10 to 45 MHz
3
27 dB 27 dB 29 dB 29 dB
45 MHz to 10 GHz 30 dB 30 dB 30 dB 30 dB
10
ALC range starts at
10 to 20 GHz 27 dB 27 dB 27 dB 27 dB
11
maximum leveled output
20 to 30 GHz 23 dB 23 dB 23 dB 23 dB power and decreases by
30 to 40 GHz 19 dB 19 dB 19 dB 19 dB power level indicated in
40 to 45 GHz 20 dB 20 dB 18 dB 16 dB the table.
45 to 50 GHz 15 dB 15 dB 12 dB 10 dB
Power resolution 0.01 dB 0.01 dB 0.01 dB 0.01 dB
1. Test port output is a specification when the source is set to port 1 and a
characteristic when the source is set to port 2.
2. Preset power.
3. Typical performance.
4. ±1.5 dB for power -23 dBm.
5. Power to which the source can be set and phase lock is assured.
6. Power level linearity is a specification when the source is set to port 1 and a
typical when the source is set to port 2.
7. Test port power is specified into nominal 50 ohms.
8. Option H11 decreases maximum power level by 1 dB.
9. Option H11 decreases maximum power level by 2 dB.
10. Option H11 decreases power level by 1 dB.
11. Option H11 decreases power level by 2 dB.