Agilent Technologies E8364C Manuale Utente Pagina 40

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 65
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 39
39
E8361C
Test port input continued
Description Specification Option 014 or Supplemental
Standard Option 014 and UNL (typ.) information
Direct receiver access
input noise floor
1
10 Hz IF bandwidth
10 to 45 MHz
2
< -106 dBm
45 to 500 MHz
4, 5
< -105 dBm Online Help also includes the
500 MHz to 2 GHz < -125.5 dBm category “Direct receiver access
2 to 10 GHz < -125 dBm input noise floor, Option 080 enabled”
10 to 24 GHz < -128 dBm
24 to 30 GHz < -117.5 dBm
30 to 40 GHz < -117 dBm Option 016 degrades
40 to 45 GHz < -115 dBm performance by 2 dB.
45 to 50 GHz < -112.5 dBm
50 to 60 GHz < -111 dBm
60 to 67 GHz < -108 dBm Option 016 degrades
67 to 70 GHz
2
< -107 dBm performance by 3 dB.
1 kHz IF bandwidth
10 to 45 MHz
2
< -86 dBm
45 to 500 MHz
4, 5
< -85 dBm
500 MHz to 2 GHz < -105.5 dBm
2 to 10 GHz < -105 dBm
10 to 24 GHz < -108 dBm
24 to 30 GHz < -97.5 dBm
30 to 40 GHz < -97 dBm Option 016 degrades
40 to 45 GHz < -95 dBm performance by 2 dB.
45 to 50 GHz < -92.5 dBm
50 to 60 GHz < -91 dBm
60 to 67 GHz < -88 dBm Option 016 degrades
67 to 70 GHz
2
< -87 dBm performance by 3 dB.
Description Specification Supplemental
Receiver compression level (measured at Test Ports)
Supplemental
Specifications Information
Standard Option 014 Option 014 and UNL Typical
10 to 45 MHz
2, 3
negligible negligible negligible negligible
45 to 500 MHz
3, 6
<0.1 dB at -9.5 dBm
7
<0.1 dB at -9.5 dBm
7
<0.1 dB at -9.5 dBm
7
<0.1 dB at +0.5 dBm
7
and <0.25 dB at -3 dBm and <0.25 dB at -3 dBm and <0.25 dB at -3 dBm and <0.25 dB at +8 dBm
500 MHz to 5 GHz <0.1 dB at -8 dBm
7
<0.1 dB at -8 dBm
7
<0.1 dB at -7 dBm
7
<0.1 dB at -4 dBm
7
and <0.25 dB at -1 dBm and <0.25 dB at -1 dBm and <0.25 dB at 0 dBm and <0.25 dB at +3 dBm
5 to 30 GHz <0.1 dB at -8.5 dBm
7
<0.1 dB at -8.5 dBm
7
<0.1 dB at -6 dBm
7
<0.1 dB at -1 dBm
7
and <0.25 dB at -2 dBm and <0.25 dB at -2 dBm and <0.25 dB at +1 dBm and <0.25 dB at +6 dBm
30 to 67 GHz < 0.1 dB at -10.5 dBm
7
< 0.1 dB at -8 dBm
7
<0.1 dB at –9.5 dBm
7
<0.1 dB at -2 dBm
7, 8
and <0.15 dB at -7 dBm and <0.15 dB at -4 dBm and <0.15 dB at -6 dBm and <0.15 dB at +2 dBm
8
67 to 70 GHz
2
<0.1 dB at -2 dBm
7, 8
and <0.15 dB at +2 dBm
8
1. Total average (rms) noise power calculated as mean value of a linear magnitude
trace expressed in dBm.
2. Typical performance.
3. Coupler roll-off will reduce compression to a negligible level below 500 MHz.
4. Noise floor may be degraded by 10 dB at particular frequencies
(multiples of 5 MHz) due to spurious receiver residuals.
5. Specified value is for worst-case noise floor at 45 MHz.
6. Specified value is for worst-case compression at 500 MHz.
7. This compression level comes from the dynamic accuracy curve with -30 dBm
reference test port power.
8. Option 016 degrades performance by 3 dB.
Vedere la pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 64 65

Commenti su questo manuale

Nessun commento