Test port input continued
Description Specification Supplemental information
Standard, 014, UNL UNL and 014
Direct receiver access input noise floor
1,2
- Option 080 enabled
4
10 Hz IF bandwidth
10 to 45 MHz
< -127 dBm < -127 dBm
45 to 500 MHz
3
< -126 dBm < -126 dBm
500 MHz to 2 GHz < -132 dBm < -132 dBm
2 to 10 GHz < -131 dBm < -131 dBm
10 to 20 GHz < -133 dBm < -133 dBm
20 to 40 GHz < -124 dBm < -124 dBm Option 016 degrades performance by 2 dB
40 to 50 GHz < -122 dBm < -122 dBm Option 016 degrades performance by 2 dB
1 kHz IF bandwidth
10 to 45 MHz
< -107 dBm < -107 dBm
45 to 500 MHz
3
< -106 dBm < -106 dBm
500 MHz to 2 GHz < -112 dBm < -112 dBm
2 to 10 GHz < -111 dBm < -111 dBm
10 to 20 GHz < -113 dBm < -113 dBm
20 to 40 GHz < -104 dBm < -104 dBm Option 016 degrades performance by 2 dB
40 to 50 GHz < -102 dBm < -102 dBm Option 016 degrades performance by 2 dB
Receiver compression level (measured at test ports)
10 MHz to 20 GHz
< 0.1 dB at -5 dBm
5
and
< 0.45 dB at +5 dBm
20 to 30 GHz
< 0.1 dB at -9.5 dBm
5
and < 0.45 dB at 0 dBm
30 to 40 GHz < 0.1 dB at -12.5 dBm
5
and < 0.45 dB at -3 dBm
40 to 50 GHz < 0.1 dB at -12.5 dBm
5
and < 0.45 dB at -3 dBm
System compression level max output power See dynamic accuracy
chart
Third order intercept – Tone spacing from 100 kHz to 5 MHz
Typical:
10 to 150 MHz
+33 dBm
150 to 300 MHz +34 dBm
300 to 500 MHz +30 dBm
500 MHz to 20 GHz +24 dBm
20 to 40 GHz +18 dBm
40 to 50 GHz +15 dBm
Third order intercept – Tone spacing from 5 MHz to 20 MHz
Typical:
10 to 500 MHz
+20 dBm
500 MHz to 20 GHz +20 dBm
20 to 40 GHz +16 dBm
40 to 50 GHz +15 dBm
Third order intercept – Tone spacing from 20 MHz to 50 MHz
Typical:
10 to 500 MHz
+26 dBm
500 MHz to 20 GHz +26 dBm
20 to 40 GHz +20 dBm
40 to 50 GHz +19 dBm
1. Total average (rms) noise power calculated as mean value of a linear magnitude
trace expressed in dBm.
2. Typical performance.
3. Noise floor may be degraded by 10 dB at particular frequencies (multiples of 5 MHz)
due to spurious receiver residuals.
4. 0 Hz offset.
5. This compression level comes from the dynamic accuracy curve with -30 dB