Test port input continued
Description Specification Option UNL Supplemental
Standard or Option 014 and Option 014 information
Third Order Intercept
1
– Tone spacing from 100 kHz to 5 MHz
Typical:
10 to 500 MHz +30 dBm
500 MHz to 24 GHz +24 dBm
24 to 40 GHz +23 dBm
40 to 50 GHz +24 dBm
50 to 67 GHz +26 dBm
Third Order Intercept
1
– Tone spacing from 5 to 20 MHz
Typical:
10 to 500 MHz
Not applicable
500 MHz to 24 GHz +20 dBm
24 to 40 GHz +20 dBm
40 to 50 GHz +22 dBm
50 to 67 GHz +24 dBm
Third Order Intercept
1
– Tone spacing from 20 to 50 MHz
Typical:
10 to 500 MHz
Not applicable
500 MHz to 24 GHz +26 dBm
24 to 40 GHz +24 dBm
40 to 50 GHz +25 dBm
50 to 67 GHz +27 dBm
System compression level – at maximum leveled output power
See Dynamic Accuracy Chart
Trace noise magnitude
10 to 45 MHz
2
< 0.150 dB rms < 0.150 dB rms
45 to 500 MHz
4, 5
< 0.010 dB rms < 0.010 dB rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.006 dB rms < 0.006 dB rms ratio measurement, nominal
24 to 67 GHz < 0.006 dB rms < 0.009 dB rms power at test port
67 to 70 GHz
2
< 0.006 dB rms < 0.009 dB rms
Trace noise magnitude
2
– Option 080 enabled
3
10 to 45 MHz
< 0.150 dB rms < 0.150 dB rms
45 to 500 MHz
4, 5
< 0.010 dB rms < 0.010 dB rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.006 dB rms < 0.006 dB rms ratio measurement, nominal
24 to 67 GHz < 0.009 dB rms < 0.012 dB rms power at test port
67 to 70 GHz < 0.009 dB rms < 0.012 dB rms
Trace noise phase
10 to 45 MHz
2
< 0.800° rms < 0.800° rms
45 to 500 MHz
5
< 0.100° rms < 0.100° rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.060° rms < 0.060° rms ratio measurement, nominal
24 to 67 GHz < 0.100° rms < 0.100° rms power at test port
67 to 70 GHz
2
< 0.100° rms < 0.100° rms
Trace noise phase
2
– Option 080 enabled
3
10 to 45 MHz
< 0.800° rms < 0.800° rms 1 kHz IF bandwidth
45 to 500 MHz
5
< 0.100° rms < 0.100° rms ratio measurement, nominal
500 MHz to 24 GHz < 0.060° rms < 0.060° rms power at test port
24 to 67 GHz < 0.100° rms < 0.100° rms
67 to 70 GHz < 0.100° rms < 0.100° rms
1. Third order intercept is a typical specification that applies while the network
analyzer receiver is in its linear range.
2. Typical performance.
3. 0 Hz offset.
4. Trace noise magnitude may be degraded to 20 mdB rms at harmonic frequencies
of the first IF (8.33 MHz) below 80 MHz.
5. Specified value is for worst-case noise floor at 45 MHz.