
41
1. Stability is defined as a ratio measurement measured at the test port.
E8361C
Test port input continued
Description Specification Supplemental
Standard Option 014 information
Reference level magnitude
Range ±500 dB ±500 dB
Resolution 0.001 dB 0.001 dB
Reference level phase
Range ±500° ±500°
Resolution 0.01° 0.01°
Stability magnitude
1
Typical ratio measurement:
Measured at the test port
10 to 45 MHz
±0.05 dB/°C
45 MHz to 20 GHz ±0.02 dB/°C
20 to 40 GHz ±0.02 dB/°C
40 to 50 GHz ±0.02 dB/°C
50 to 70 GHz ±0.04 dB/°C
Stability phase
1
Typical ratio measurement:
Measured at the test port
10 to 45 MHz
±0.5°/°C
45 MHz to 20 GHz ±0.2°/°C
20 to 40 GHz ±0.5°/°C
40 to 50 GHz ±0.8°/°C
50 to 70 GHz ±0.8°/°C
Damage input level
Test port 1 and 2 +27 dBm or ±40 VDC, typical
R1, R2 in +15 dBm or ±15 VDC, typical
A, B in +15 dBm or ±7 VDC, typical
Coupler thru (Option 014) +27 dBm or ±40 VDC, typical
Coupler arm (Option 014) +30 dBm or ±7 VDC, typical
Source out (reference) +15 dBm or ±15 VDC, typical
Source out (test ports) +27 dBm or ±5 VDC, typical
Commenti su questo manuale