49
1. Measured at –5 dBm test port power.
2. Pulse input connectors are operational only with Option H08 (Pulsed Measurement Capability) enabled.
3. Based on deviation from signal reduction equation:
Signal Reduction (dB) = 20log
10
(Duty_cycle) = 20log
10
(pulse_width/pulse_repetition_interval).
Measured at pulse repetition frequency of 1 MHz.
Microwave PNA Series
Rear panel
Description Supplemental information
External trigger rear panel I/O (typical)
Trigger inputs/outputs BNC(f), TTL/CMOS compatible
Trigger input
Function Measurement of next point, next channel, or next group of channels
Source Aux I/O (pin 19) or I/O 1 (BNC (f) connector)
Signal levels TTL-compatible
Input impedance 5 k nominal
Minimum trigger width 1 µs
Trigger modes High or low level; positive or negative edge
Trigger delay range 0 to 1 sec
Trigger delay resolution 6 µs (IF bandwidth 15 kHz) or 6.2 us (IF bandwidth <15 kHz)
Trigger output
Function Generate pulse before or after measurement
(only active when trigger type is external)
Source I/O 2 (BNC (f) connector)
Signal levels TTL-compatible
Trigger polarity Positive or negative edge
Pulse width 1 µs
Option H11 rear panel I/O (typical)
External IF inputs
Function Allows use of external IF signals from remote mixers, bypassing the
PNA’s first converters
Connectors BNC (f), for B, R2, R1, A receivers
Input frequency 8.33 MHz
Input impedance 50 nominal
RF damage level –20 dBm
DC damage level 25 Volts
0.1 dB compression point –27 dBm
Test Set Drivers
Function Used for driving remote mixers
Connectors SMA (f) for RF and LO outputs
RF, LO output frequency range 1.7 to 20 GHz
RF output power levels +5 to –16 dBm, depending on frequency
1
LO output power levels –7 to –16 dBm, depending on frequency
Pulse inputs (IF gates)
2
Function Internal receiver gates used for point-in-pulse and pulse-profile
measurements
Connectors BNC (f), for B, R2, R1, A receivers
Input impedance 1 k nominal
Minimum pulse width 20 ns for less than 1 dB deviation from theoretical performance
3
DC damage level 5.5 Volts
Signal levels TTL; 0 V (off), +5 V (on) nominal