62
Test port cable specifications
Connector Type Frequency Length
2
Return Insertion loss Stabillty
1,2
±Phase
(Test port to device) (GHz) cm (inch) loss (dB) (f in GHz) ±magnitude (degrees)
Single cables
(3.5 mm)
85131C semi-rigid cable 3.5 mm
3
to DC to 26.5 81 (32) 17 dB 0.43 f +0.3 <0.06 dB 0.16 (f) +0.5
PSC-3.5 mm (f) (2.5 dB at f
max
)
85131E flexible cable 3.5 mm
3
to DC to 26.5 96.5 (38) 16 dB 0.35 f +0.3 <0.22 dB 0.16 (f) +0.8
PSC-3.5 mm (f) (2.1 dB at f
max
)
85132C semi-rigid cable 3.5 mm
3
to 7 mm DC to 18 81 (32) 17 dB 0.35 f +0.3 <0.06 dB 0.16 (f) +0.5
(1.8 dB at f
max
)
85132E flexible cable 3.5 mm
3
to 7 mm DC to 18 97.2 (38.25) 17 dB 0 35 f +0.3 <0.22 dB 0.16 (f) +0.8
(1.8 dB at f
max
)
Cable sets
(3.5 mm)
85131D semi-rigid cable set 3.5mm
3
to DC to 26.5 53 (21) 16 dB 0.30 f +0.2 <0.06 dB 0.16 (f) +0.5
PSC-3.5 mm (f) or (1.8 dB at f
max
)
3.5 mm (m)
85131F flexible cable set 3.5 mm
3
to DC to 26.5 53 (21) 16 dB 0.25 f +0.2 <0.12 dB 0.13 (f) +0.5
PSC-3.5 mm (f) or (1.5 dB at f
max
)
3.5 mm (m)
85132D semi-rigid cable set 3.5 mm
3
to 7 mm DC to 18 53 (21 ) 17 dB 0.25 f +0.2 <0.06 dB 0.16 (f) +0.5
(1.3 dB at f
max
)
85132F flexible cable set 3.5 mm
3
to 7 mm DC to 18 53 (21) 17 dB 0.25 f +0.2 <0.12 dB 0.13 (f) +0.5
(1.3 dB at f
max
)
N4419A-B20 flexible cable set 3.5 mm(m) to DC to 26.5 91.4 (36) >/=15 -1.9 dB at 26.5 GHz <0.15 dB 6.56
3.5 mm (f)
Single cables
(2.4 mm)
85133C semi-rigid cable 2.4 mm
3
to DC to 50 81 (32) 15 dB 0.84 f +0.3 <0.06 dB 0.18 (f)
PSC-2.4 mm (f) (5.6 dB at f
max
)
85133E flexible cable 2.4 mm
3
to DC to 50 113 (44) 12.5 dB 0.58 f +0.35 <0.25 dB 0.8 +0.16 (f)
PSC-2.4 mm (f) (4.45 dB at f
max
)
85134C semi-rigid cable 2.4 mm
3
to DC to 26.5 81 (32) 16 dB 0.46 f +0.3 <0.06 dB 0.18 (f)
PSC-3.5 mm (f) (2.7 dB at f
max
)
85134E flexible cable 2.4 mm
3
to DC to 26.5 97.2 (38.25) 16 dB 0.46 f +0.3 <0.22 dB 0.16 (f) +0.8
PSC-3.5 mm (f) (2.7 dB at f
max
)
85135C semi-rigid cable 2.4 mm
3
to 7 mm DC to 18 81 (32) 17 dB 0.46 f +0.3 <0.06 dB 0.18 (f)
(2.25 dB at f
max
)
85135E flexible cable 2.4 mm
3
to 7 mm DC to 18 97.2 (38.25) 17 dB 0.46 f +0.3 <0.22 dB 0.16 (f) +0.8
(2.25 dB at f
max
)
1. Phase stability of semi-rigid/flexible cables is specified with a 90-degree bend and a 4"/3" radius.
2. Cable length and stability are supplemental characteristics.
3. Special rugged female connector specifically for connecting to the network analyzer test port. Does not mate with a standard male connector.
Commenti su questo manuale