Contents 103 Index
6. Two proteins are run with each of the samples, bracketing the sizing range. The “lower
marker” and “upper marker” proteins are internal standards used to align the ladder
data with data from the sample wells. The figure below shows an example of assigned
marker peaks in a sample well.
NOTE
The software performs alignment by default. Turning automatic data analysis off (see
Electropherogram Menu” on page 345) suspends analysis until you turn it on again.
Lower
marker
Upper marker
Vedere la pagina 102