Agilent Technologies DP235 Guida Utente Pagina 52

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 602
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 51
Contents 52 Index
How to Prepare the Bioanalyzer for Flow Cytometric Assays
Remove the electrode cartridge:
1 Turn off line power to the Agilent 2100 bioanalyzer. The line switch is located at the rear.
2 Open the lid.
CAUTION
Do not touch the electrodes while the cartridge is in the Agilent 2100 bioanalyzer. The
electrodes and the high voltage power supplies can be damaged.
3 Pull down the metal locking lever in the open position as shown in the figure below.
The cartridge is pushed out.
4 Gently pull the cartridge out of the lid.
Metal lever
in open
position
Vedere la pagina 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 601 602

Commenti su questo manuale

Nessun commento