230 Chapter 2
Performance Verification Tests
26. Frequency Response (Preamp On): Agilent E4402B
9. Note the most negative number from the Flatness Relative to
50MHz column of Table 2-61 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-62 as the Minimum
Response for Band 0B.
10.For Band 0B in Table 2-62, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
11.Transfer the values entered in Table 2-62 to the test record in this
guide.
12.The frequency response test is now complete.
Table 2-62 Frequency Response Results
Band
Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
01) 2) 3)
0A 4) 5) 6)
0B 7) 8) 9)