Chapter 2 303
Performance Verification Tests
35. Displayed Average Noise Level: Agilent E4401B and E4411B
perform those procedures listed in Subset C in the appropriate
table.
7. If the minimum RBW of the analyzer is 10 Hz and Option 1DS (RF
Preamplifier) is installed, also perform those procedures listed in
Subset D in the appropriate table.
8. After performing all applicable DANL measurement procedures,
continue with “Remove Reference Level Offset.