374 Chapter 2
Performance Verification Tests
48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B (Option 1DN)
start and stop frequency settings.
16.Repeat step 8 through step 15 for the remaining analyzer center
frequency and microwave analyzer settings in Table 2-104.
17.Press
System, Alignments, Auto Align, All
Determining the Highest Non-harmonic Spurious Response
1. In Table 2-104, locate the most positive non-harmonic response
amplitude for microwave analyzer stop frequency settings
2000 MHz. Record this amplitude as the highest non-harmonic
response amplitude in Test Record entry 1 of the performance
verification test record.
2. In Table 2-104, locate the most positive non-harmonic response
Table 2-104 3.0 GHz Tracking Generator Non-Harmonic Spurious Response
Worksheet
Analyzer
Center
Frequency
Microwave
Analyzer Start
Frequency
Microwave
Analyzer Stop
Frequency
Microwave
Analyzer
Resolution
Bandwidth
Amplitude of
Non-Harmonic
Response
(dBc)
10 MHz 9 kHz 100 kHz 300 Hz
10 MHz 100 kHz 5 MHz 10 kHz
10 MHz 5 MHz 55 MHz 100 kHz
10 MHz 55 MHz 1240 MHz 1 MHz
10 MHz 1240 MHz 2000 MHz 1 MHz
10 MHz 2000 MHz 3000 MHz 1 MHz
1.5 GHz 9 kHz 100 kHz 300 Hz
1.5 GHz 100 kHz 5 MHz 10 kHz
1.5 GHz 5 MHz 55 MHz 100 kHz
1.5 GHz 55 MHz 1240 MHz 1 MHz
1.5 GHz 1240 MHz 2000 MHz 1 MHz
1.5 GHz 2000 MHz 3000 MHz 1 MHz
3.0 GHz 9 kHz 100 kHz 300 Hz
3.0 GHz 100 kHz 5 MHz 10 kHz
3.0 GHz 5 MHz 55 MHz 100 kHz
3.0 GHz 55 MHz 1240 MHz 1 MHz
3.0 GHz 1240 MHz 2000 MHz 1 MHz
3.0 GHz 2000 MHz 3000 MHz 1 MHz