Chapter 3 577
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
ac Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (28)________ 1.5 dB
±0.39 dB
Minimum Response –1.5 dB (29)________
±0.39 dB
Peak-to-Peak Response (30)________ 2.6 dB ±0.39 dB
ac Coupled
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (31)________ 2.0 dB
±0.40 dB
Minimum Response –2.0 dB (32)________
±0.40 dB
Peak-to-Peak Response (33)________ 3.6 dB
±0.40 dB
ac Coupled
Band 3
(13.2 GHz to 26.5 GHz)
Maximum Response (34)________ 2.0 dB
±0.59 dB
Minimum Response –2.0 dB (35)________ ±0.59 dB
Peak-to-Peak Response (36)________ 3.6 dB
±0.59 dB
0 to 55
° C
Non-Option UKB
dc Coupled
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.76 dB
±0.19 dB
Minimum Response –0.76 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 1.52 dB ±0.19 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Vedere la pagina 576
1 2 ... 572 573 574 575 576 577 578 579 580 581 582 ... 641 642