440 Chapter 3
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
200 Hz (Option 1DR only) –0.3 dB (12)________ 0.3 dB ±0.048 dB
100 Hz (Option 1DR only) –0.3 dB (13)________ 0.3 dB
±0.048 dB
30 Hz (Option 1DR only) –0.3 dB (14)________ 0.3 dB
±0.048 dB
10 Hz (Option 1DR only) –0.3 dB (15)________ 0.3 dB ±0.048 dB
3 Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (16)________ 0.3 dB
±0.048 dB
1Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (17)________ 0.3 dB
±0.048 dB
17. Absolute Amplitude
Accuracy (Reference
Settings)
Log, Preamp Off –0.30 dB (1)________ 0.30 dB
±0.09 dB
Lin, Preamp Off –0.30 dB (2)________ 0.30 dB
±0.09 dB
Log, Preamp On
(Option 1DS only)
–0.37 dB (3)________ 0.37 dB
±0.09 dB
Lin, Preamp On
(Option 1DS only)
–0.37 dB (4)________ 0.37 dB
±0.09 dB
19. Overall Absolute
Amplitude Accuracy
0 dBm Reference Level
0 dBm input –0.54 dB (1)________ 0.54 dB
±0.234 dB
–10 dBm input –0.54 dB (2)________ 0.54 dB
±0.115 dB
–20 dBm input –0.54 dB (3)________ 0.54 dB ±0.110 dB
–30 dBm input –0.54 dB (4)________ 0.54 dB
±0.109 dB
–40 dBm input –0.54 dB (5)________ 0.54 dB
±0.109 dB
–50 dBm input –0.54 dB (6)________ 0.54 dB ±0.109 dB
–20 dBm Reference Level
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Vedere la pagina 439