380 Chapter 2
Performance Verification Tests
50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,
E4405B, and E4407B (Option 1D6)
Figure 2-70 Gate Delay and Gate Length Test Setup
3. Set up the function generator as follows:
4. Press
Default Setup on the oscilloscope and set the controls as follows:
Waveform Square
Frequency 100 Hz
Amplitude 2.5 V
Offset 1.25 V
Timebase 500 ns/div
Trigger Edge
Trigger Source 2
Trigger Level 2.0 V
Channel 1
V/Div 1 V
Coupling DC
Offset 2.0 V
Channel 2
V/Div 1 V
Coupling DC