Agilent Technologies E1401-90493 Manuale Utente Pagina 186

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 642
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 185
186 Chapter 2
Performance Verification Tests
23. Frequency Response, Agilent E4402B and E4403B
and 2.0 GHz. Record this number in Table 2-48 as the Minimum
Response for Band 0B.
26.For Band 0B in Table 2-48, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
27.Transfer the values entered in Table 2-48 to the test record in this
guide.
28.The frequency response test for analyzers with Option UKB is now
complete.
Table 2-47 Frequency Response Results, dc Coupled
Band Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
01) 2) 3)
0A 4) 5) 6)
0B 7) 8) 9)
Table 2-48 Frequency Response Results, ac Coupled
Band
Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
0 10) 11) 12)
0A 13) 14) 15)
0B 16) 17) 18)
Vedere la pagina 185
1 2 ... 181 182 183 184 185 186 187 188 189 190 191 ... 641 642

Commenti su questo manuale

Nessun commento