Chapter 3 585
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Option UKB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (16)________ 2.0 dB
±0.28 dB
Minimum Response –2.0 dB (17)________
±0.28 dB
Peak-to-Peak Response (18)________ 4.0 dB
±0.28 dB
29. Other Input-Related
Spurious Responses
Center Freq Input Freq
2.0 GHz 2042.8 MHz (1)________ –65 dBc
±.891 dB
2.0 GHz 2642.8 MHz (2)________ –65 dBc
±.891 dB
2.0 GHz 1820.8 MHz (3)________ –65 dBc
±.891 dB
2.0 GHz 278.5 MHz (4)________ 65 dBc
±.891 dB
2.0 GHz 5600.0 MHz (5)________ –80 dBc
±.891 dB
2.0 GHz 6242.8 MHz (6)________ –80 dBc
±.891 dB
4.0 GHz 4042.8 MHz (7)________ –65 dBc ±.891 dB
4.0 GHz 4642.8 MHz (8)________ –65 dBc
±.891 dB
4.0 GHz 3742.9 MHz (9)________ –65 dBc
±.891 dB
4.0 GHz 2242.8 MHz (10)________ –80 dBc ±.891 dB
9.0 GHz 9042.8 MHz (11)________ –65 dBc
±.891 dB
9.0 GHz 9642.8 MHz (12)________ –65 dBc
±.891 dB
9.0 GHz 9342.8 MHz (13)________ –65 dBc ±.891 dB
9.0 GHz 4982.1 MHz (14)________ –80 dBc
±.891 dB
15.0 GHz 15042.8 MHz (15)________ –65 dBc
±.891 dB
15.0 GHz 15642.8 MHz (16)________ –65 dBc ±.891 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty