214 Chapter 2
Performance Verification Tests
25. Frequency Response (Preamp On): Agilent E4401B
Procedure
Source/Splitter Characterization
1. Refer to Figure 2-36. Connect one RF power sensor to Channel A of
the power meter. This will be the “reference” sensor. Connect the
other RF power sensor to Channel B of the power meter. This will be
the “buried” sensor.
75
inputs: Connect the 75 power sensor to Channel A of the power
meter. This will be the “reference” sensor.
Figure 2-36 Source/Splitter Characterization Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or damage to the connectors will occur.
2. Zero and calibrate both power sensors.
3. On the power meter, set the Channel A calibration factor to the
calibration factor of the reference sensor at 100 kHz.
75
inputs: Set the Channel A calibration factor to the calibration
factor of the reference sensor at 1 MHz.
4. On the power meter, set the Channel B calibration factor to 100%.
Do not change this calibration factor during this test.
5. Connect the equipment as shown in Figure 2-36. Note that the
reference sensor connects to the 20 dB fixed attenuator.
Vedere la pagina 213