Chapter 2 149
Performance Verification Tests
19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B
c. Press Single and wait for the sweep to finish.
d. Press
Peak Search (or Search). Even though the signal may be
slightly above the reference level for the first nominal amplitude
setting, the marker can still make a valid measurement.
e. Record the marker (Mkr1) amplitude value as the measured
amplitude in Table 2-33.
f. Calculate the amplitude accuracy as follows:
g. Record the amplitude accuracy in the performance verification
test record as indicated in Table 2-33.
Amplitude Accuracy Measured Amplitude Amptd
0dBm
ActualTotalAtten RefAtten
0dB
+()=