70 Chapter 2
Performance Verification Tests
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
8. Noise Sidebands - Wide Offsets: Agilent
E4402B, E4404B, E4405B and E4407B
A 1 GHz CW signal is applied to the input of the analyzer. The marker
functions are used to measure the amplitude of the carrier and the
noise level at 100 kHz, 1 MHz, 5 MHz, and 10 MHz above and below the
carrier.
There are no related adjustment procedures for this performance test.
Equipment Required
Wide offset phase noise signal generator
Cable, Type-N, 152-cm (60-in)
Additional Equipment for Option BAB
Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-8 Noise Sidebands Test Setup
Procedure
This performance test consists of four parts:
Part 1: Noise Sideband Suppression at 100 kHz
Part 2: Noise Sideband Suppression at 1 MHz
Part 3: Noise Sideband Suppression at 5 MHz
Part 4: Noise Sideband Suppression at 10 MHz
Perform part 1 before performing parts 2-4 of this procedure.
A worksheet is provided at the end of this procedure for calculating the