Chapter 2 75
Performance Verification Tests
9. System-Related Sidebands
9. System-Related Sidebands
A 500 MHz CW signal is applied to the input of the analyzer. The
marker functions are used to measure the amplitude of the carrier and
the amplitude of any system-related sidebands more than 30 kHz away
from the carrier. System-related sidebands are any internally generated
sidebands related to the line, power supply or local oscillator.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized signal generator
Cable, Type-N, 152-cm (60-in)
Additional Equipment for 75 Input
Pad, minimum loss
Adapter, Type-N (f), to BNC (m), 75
Additional Equipment for Option BAB
Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-9 System-related Sidebands Test Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or the connectors will be damaged.
Procedure
1. Perform the following steps to set up the equipment: