Calibration GuideAgilent Technologies ESA Spectrum Analyzers This manual provides documentation for the following instruments:Agilent ESA-E SeriesE44
10 Chapter 1 Calibrating Calibration verifies that the analyzer performance is within all specifications. It is time consuming and requires extensi
100 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertainty13. Input Attenuation Switching UncertaintyA 50 MHz CW signal i
Chapter 2 101Performance Verification Tests13. Input Attenuation Switching UncertaintyFigure 2-13 Input Attenuator Switching Uncertainty Test Setup C
102 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertaintyusing the Agilent 8496G programmable attenuator, enter the cali
Chapter 2 103Performance Verification Tests13. Input Attenuation Switching Uncertainty4. For each attenuation error value in Column 4 of Table 2-19, c
104 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertainty7. Preset the synthesized signal generator (Blue Key, Special,
Chapter 2 105Performance Verification Tests13. Input Attenuation Switching UncertaintyMeasure Switching UncertaintyPerform step 12 to step 15 for each
106 Chapter 2 Performance Verification Tests13. Input Attenuation Switching UncertaintyExample for 25 dB analyzer internal attenuation setting:Post-te
Chapter 2 107Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411B14. Reference Level Accuracy: Agilent E4401B and E44
108 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BProcedure Figure 2-14 Reference Level Accuracy Tes
Chapter 2 109Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BExample for 50 dB attenuator setting:Log Scale, Anal
Chapter 1 11 CalibratingCalibration CycleCalibration Cycle The performance tests in chapter 2 should be used to check the analyzer against its spec
110 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Bamplitude and enter the result as the Test Record e
Chapter 2 111Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLog Scale, Digital Bandwidths, Option 1DRNOTE If the
112 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Ba. Press Single on the analyzer.b. Press Peak Searc
Chapter 2 113Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLinear Scale, Analog Bandwidths1. Set the 10 dB step
114 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BPeak Search (or Search)Marker, Delta6. Set the 10 d
Chapter 2 115Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLinear Scale, Digital Bandwidths, Option 1DRNOTE If
116 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Ba. Press Single on the analyzer.b. Press Peak Searc
Chapter 2 117Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.15. Reference Level
118 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.Figure 2-15 Refe
Chapter 2 119Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.Example for 50 dB a
12 Chapter 1 Calibrating Calibration CycleTable 1-1 Agilent E4401B Performance Verification TestsPerformance Test NameCalibration for Instrument O
120 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.performance test
Chapter 2 121Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.9. On the analyzer,
122 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.AMPLITUDE, Scale
Chapter 2 123Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.Linear Scale, Digit
124 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.b. Press Peak Sea
Chapter 2 125Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty16. Resolution Bandwidth Switching UncertaintyTo measure the
126 Chapter 2 Performance Verification Tests16. Resolution Bandwidth Switching UncertaintyInput/Output (or Input), Amptd Ref (On) (Agilent E4401B, E44
Chapter 2 127Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty7. If you are testing an analyzer equipped with Option 1DR, p
128 Chapter 2 Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty10 Hzb100 Hz 15)3 Hzc,b100 Hz 16)1 Hzc,b100 Hz 17)a. These R
Chapter 2 129Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B17. Absolute Amplitude Accur
Chapter 1 13 CalibratingCalibration CycleTable 1-2 Agilent E4402B Performance Verification TestsPerformance Test NameCalibration for Instrument Op
130 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411Bequipped with Option 1DS, P
Chapter 2 131Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B7. Adjust the synthesized si
132 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BFREQUENCY, 50 MHzAMPLITUDE,
Chapter 2 133Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BBW/ Avg, Resolution BW, 1 kH
134 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BAbsolute Amplitude Accuracy
Chapter 2 135Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BDet/Demod, Detector, Sample6
136 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 137Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
138 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 139Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
14 Chapter 1 Calibrating Calibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Com
140 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 141Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
142 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 143Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B19. Overall Absolute Amplitude Accuracy:
144 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B3. Zero and calibrate the power meter an
Chapter 2 145Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B8. Obtain the metrology data for the step
146 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411BFigure 2-22 Amplitude Accuracy Test Set
Chapter 2 147Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411Btest record as indicated in Table 2-30.Me
148 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B3. Perform the following steps for each
Chapter 2 149Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411Bc. Press Single and wait for the sweep to
Chapter 1 15 CalibratingCalibration CycleTable 1-3 Agilent E4403B Performance Verification TestsPerformance Test NameCalibration for Instrument Op
150 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B20. Ove
Chapter 2 151Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B4. Zero a
152 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Bbe used
Chapter 2 153Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408BFigure 2-
154 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BMeasuri
Chapter 2 155Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408BTable 2-3
156 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Ba. Set
Chapter 2 157Performance Verification Tests21. Resolution Bandwidth Accuracy21. Resolution Bandwidth AccuracyThe output of a synthesized signal genera
158 Chapter 2 Performance Verification Tests21. Resolution Bandwidth AccuracyFigure 2-25 Resolution Bandwidth Accuracy Test Setup CAUTION Use only 75
Chapter 2 159Performance Verification Tests21. Resolution Bandwidth AccuracyAttenuator Error (6 dB) ___________ dB3 dB Resolution Bandwidth Accuracy6.
16 Chapter 1 Calibrating Calibration CycleTable 1-4 Agilent E4404B Performance Verification TestsPerformance Test NameCalibration for Instrument O
160 Chapter 2 Performance Verification Tests21. Resolution Bandwidth AccuracyFrequency6 dB Resolution Bandwidth Accuracy18.Set the analyzer Res BW to
Chapter 2 161Performance Verification Tests21. Resolution Bandwidth Accuracy28.Press Marker, Normal on the analyzer. 29.Repeat step 19 through step 28
162 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B22. Frequency Response: Agilent E4401B and E4411BThis tes
Chapter 2 163Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BAdditional Equipment for 75 Ω InputPower sensor, 75 ΩMinim
164 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411Bcalibration factor of the reference sensor for 100 kHz.75
Chapter 2 165Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BTable 2-40 Source/Splitter CharacterizationFrequencyPower
166 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BMeasuring Frequency Response, 100 kHz to 1.5 GHz1. Refer
Chapter 2 167Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BVideo BW, 3 kHz (Man)5. Adjust the source AMPLITUDE to obt
168 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B0.15 dB, Flatness Relative to 50 MHz would be –0.45 dB.Ta
Chapter 2 169Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BMeasuring Frequency Response, ≤100 kHz If the analyzer has
Chapter 1 17 CalibratingCalibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Comm
170 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B10.On the analyzer, press Peak Search (or Search).11.Adju
Chapter 2 171Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B4. Enter the most negative number from the Flatness Relati
172 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B23. Frequency Response, Agilent E4402B and E4403BThis tes
Chapter 2 173Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BProcedureSource/Splitter Characterization1. Connect the eq
174 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B10.Adjust the source amplitude to obtain a Channel A powe
Chapter 2 175Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B475 MHz575 MHz675 MHz775 MHz825 MHz875 MHz925 MHz975 MHz10
176 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BMeasuring Frequency Response, 100 kHz to 3.0 GHz1. Refer
Chapter 2 177Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BCF Step, 100 MHzSPAN, 20 kHzInput/Output (or Input), Coupl
178 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403Bas marker (Mkr1) Amplitude.24.Repeat step 18 through step
Chapter 2 179Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BFor example, if Flatness Error is –0.30 dB and 50 MHz Ref
18 Chapter 1 Calibrating Calibration CycleTable 1-5 Agilent E4405B Performance Verification TestsPerformance Test NameCalibration for Instrument O
180 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B575 675 775 825 875 925 975 1025 1075 1175 1275 1375 1500
Chapter 2 181Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B2925 2999 a. This entry is for data taken with the functio
182 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BMeasuring Frequency Response Less Than or Equal to 100 kH
Chapter 2 183Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BTable 2-46. Frequencies less than 9 kHz only apply to anal
184 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BTest Results Perform the following steps to verify the fr
Chapter 2 185Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B13.Note the most positive number from the Flatness Relativ
186 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403Band 2.0 GHz. Record this number in Table 2-48 as the Mini
Chapter 2 187Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B24. Frequency Response, Agilent E4404B, E
188 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAdapter, APC 3.5 (f) to APC 3.5 (f) Adap
Chapter 2 189Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408Bpower meter reading of –10 dBm ±0.1 dB.7.
Chapter 1 19 CalibratingCalibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Comm
190 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B20 MHz 0 dB (Ref)50 MHz75 MHz175 MHz275
Chapter 2 191Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response, 100 kHz to
192 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B2. Set the source frequency to 10 MHz.3.
Chapter 2 193Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B18.Set the function generator amplitude t
194 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BSetup Change Error =__________ dB30.For
Chapter 2 195Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B10c 20 50 0 dB (ref)0 dB (ref)75 175 275
196 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B1825 1875 1925 1975 2025 2125 2325 2525
Chapter 2 197Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response Less Than or
198 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B9. On the analyzer, press Peak Search (o
Chapter 2 199Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B500 Hza200 Hza100 Hzaa. These frequencies
2 NoticeThe information contained in this document is subject to change without notice.Agilent Technologies makes no warranty of any kind with regar
20 Chapter 1 Calibrating Calibration CycleTable 1-6 Agilent E4407B Performance Verification TestsPerformance Test NameCalibration for Instrument O
200 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response, >3 GHz1
Chapter 2 201Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BPresel Center.13.Adjust the synthesized s
202 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B26.Repeat step 10 through step 14 for fr
Chapter 2 203Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B5.25 GHz5.5 GHz5.75 GHz6.0 GHz6.25 GHz6.5
204 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B13.5 GHz14.0 GHz14.5 GHz15.0 GHz15.5 GHz
Chapter 2 205Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAgilent E4404B/E4405B Test Results Perfor
206 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B50 MHz (dc) column of Table 2-52 for fre
Chapter 2 207Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B6.7 GHz. Record this number in Table 2-54
208 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAgilent E4407B/E4408B Test Results Perfo
Chapter 2 209Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408Bcontinue with step 15.9. Note the most po
Chapter 1 21 CalibratingCalibration Cycle52. First LO OUTPUT Power Accuracy •53. IF Input Accuracy •54. Comms Frequency Response ••55.(This test ha
210 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B20.For Band 2 in Table 2-55, subtract th
Chapter 2 211Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMHz (ac) column of Table 2-50. Record thi
212 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B14.Note the most negative number from th
Chapter 2 213Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B25. Frequency Response (Preamp On): Agilent E4401BThis te
214 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BProcedureSource/Splitter Characterization1. Refer to Fi
Chapter 2 215Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B75 Ω inputs: Connect the reference sensor to the power sp
216 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BMeasuring Frequency Response, Preamp On1. Refer to Figur
Chapter 2 217Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BFigure 2-37 Frequency Response Test Setup, Preamp On75 Ω
218 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B8. Record the marker (Mkr1) amplitude reading in Table 2
Chapter 2 219Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BTest Results1. Record the most positive number from the F
22 Chapter 1 Calibrating Calibration CycleTable 1-7 Agilent E4408B Performance Verification TestsPerformance Test NameCalibration for Instrument O
220 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B3. In Table 2-59, subtract the Minimum Response value fr
Chapter 2 221Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B26. Frequency Response (Preamp On): Agilent E4402BThis te
222 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Bcalibration factor of the reference sensor at 1 MHz.4. O
Chapter 2 223Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B15.Set the synthesized sweeper CW frequency to 10 MHz and
224 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402BMeasuring Frequency Response, Preamp On1. Refer to Figur
Chapter 2 225Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Buse a cable. 2. Set the source frequency to 10 MHz.3. Set
226 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B10.Set the analyzer center frequency to the next frequen
Chapter 2 227Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Bgenerator) as follows:a. Record the Flatness Error from T
228 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B175 MHz275 MHz375 MHz475 MHz575 MHz675 MHz775 MHz825 MHz
Chapter 2 229Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402BTest Results1. Record the most positive number from the F
Chapter 1 23 CalibratingCalibration CycleTable 1-8 Agilent E4411B Performance Verification TestsPerformance Test NameCalibration for Instrument Op
230 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B9. Note the most negative number from the Flatness Relat
Chapter 2 231Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B27. Frequency Response (Preamp On): A
232 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BChannel B of the power meter. This w
Chapter 2 233Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTable 2-63.13.Repeat step 9 through s
234 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B775 MHz825 MHz875 MHz925 MHz975 MHz1
Chapter 2 235Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BMeasuring Frequency Response, Preamp
236 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B5. On the analyzer, press Peak Searc
Chapter 2 237Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B27.Calculate the Flatness Error for e
238 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B33.If the analyzer is an Agilent E44
Chapter 2 239Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTable 2-64 Frequency Response Worksh
24 Chapter 1 Calibrating Before You StartBefore You Start This brief procedure should be performed before starting the performance verification tes
240 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTest Results1. Record the most posit
Chapter 2 241Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B5. If the analyzer is an Agilent E440
242 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B15.The frequency response (preamp on
Chapter 2 243Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B25.Transfer the values entered in Tab
244 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B28. Other Input-Related Spurious Resp
Chapter 2 245Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B75 Ω Input only: Use the minimum loss
246 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B16.On the synthesized sweeper, press
Chapter 2 247Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B24.On the synthesized sweeper, press R
248 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411BTable 2-67 Other Input-Related Spuri
Chapter 2 249Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B29. Ot
Chapter 1 25 CalibratingBefore You StartPeriodic verification of operation The analyzer requires periodic verification of operation. Under most con
250 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BFigu
Chapter 2 251Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B10.Rep
252 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Bfreq
Chapter 2 253Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B30.On
254 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B15.0
Chapter 2 255Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B30. Spurious Responses: Agilent E4401B and E4411BThis test
256 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411BAdapter, Type-N (m) to BNC (f) Adapter, Type-N (m) to Typ
Chapter 2 257Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411BFigure 2-45 Third Order Intermodulation Distortion Test S
258 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411Bfrequency into the power meter. 7. Adjust the amplitude o
Chapter 2 259Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B17.Set the analyzer reference level to –15 dBm. 75 Ω Input
26 Chapter 1 Calibrating Before You StartOscilloscope Bandwidth: dc to 100 MHz Vertical Scale Factor of 0.5 V to 5 V/DivTwo channelsMinimum Timebas
260 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B30.If the analyzer is an E4401B and is equipped with Opti
Chapter 2 261Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B36.Repeat step 10 through step 29. This is the TOI test fo
262 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B5. Set the analyzer control as follows: SPAN, 50 kHzBW/Av
Chapter 2 263Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B31. Spurious Responses: Agilent E4402B and E4403BThis test
264 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BAdapter, Type-N (m) to BNC (f) Adapter, Type-N (m) to Typ
Chapter 2 265Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B3. Perform step 4 through step 29 using the information an
266 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BMarker →Mkr → CFDelta11.On the analyzer, press: FREQUENCY
Chapter 2 267Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BDistortion Amplitude (in dB) from the Mixer Level (in dBm)
268 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B36.Repeat step 10 through step 29. This is the TOI test f
Chapter 2 269Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B6. Record the power meter reading at 300 MHz in Table 2-73
Chapter 1 27 CalibratingBefore You StartWide Offset Phase Noise Signal GeneratorFrequency Range: 1 GHz ±1 MHzAmplitude Range: 0 dBM ±5 dBPhase Nois
270 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B17.Record the power meter reading in Table 2-73 as the 90
Chapter 2 271Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B5. Adjust the synthesized sweeper power level until the po
272 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BAMPLITUDE, Ref Level, –20 dBmBW/Avg, Video BW, 30 Hz (Man
Chapter 2 273Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B32. Spurious Responses: Agilent E4404B, E
274 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B1.8 GHz low pass filter (2 required)4.4
Chapter 2 275Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BFigure 2-49 Third Order Intermodulation
276 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bmeter reads –12 dBm ±0.1 dB.8. Disconnec
Chapter 2 277Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B20.On the analyzer, press BW/Avg, Average
278 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B31.If the analyzer is an E4404B, E4405B
Chapter 2 279Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BPart 2: Second Harmonic Distortion 1. Zer
28 Chapter 1 Calibrating Before You StartAttenuator, 1 dB StepAttenuation Range: 0 to 11 dB Frequency Range: 50 MHz ±1 MHzConnectors: Type-N female
280 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BFigure 2-50 Second Harmonic Distortion
Chapter 2 281Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B11.Subtract the power meter reading at 60
282 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bat 900 MHz is –7.05 dBm, the 900 MHz Fre
Chapter 2 283Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bwould be –1.2 dB:Measuring 3.1 GHz Freque
284 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bbe –1.2 dB:Measuring 300 MHz Second Harm
Chapter 2 285Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Breading is –10 dBm ±0.2 dB.6. On the anal
286 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BAMPLITUDE, Ref Level, –20 dBmBW/Avg, Vid
Chapter 2 287Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BMeasuring 1.55 GHz Second Harmonic Distor
288 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B13.Calculate the 1.55 GHz Second Harmoni
Chapter 2 289Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bthe synthesized sweeper output directly t
Chapter 1 29 CalibratingBefore You StartPower Splitter(for E4401B/ 02B/03B/11B)Frequency Range: 9 kHz to 13.2 GHz Insertion Loss: 6 dB nominal Outp
290 Chapter 2 Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B33. Gain Compression: Agilent E4401B, E440
Chapter 2 291Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411Bdirectional bridge connected to the power s
292 Chapter 2 Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B7. Record the actual synthesized sweeper p
Chapter 2 293Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B10.Set the synthesized sweeper amplitude Of
294 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B34. Gain Compression: Agilent E4404B, E440
Chapter 2 295Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408Bless than or equal to 2.5 GHz. Use the dire
296 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408BTable 2-80 for each frequency indicated.Ta
Chapter 2 297Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B8. Repeat step 3 through step 7 for each of
298 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B21.Replace the directional bridge with the
Chapter 2 299Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411B35. Displayed Average Noise Level: Agilent E440
3Where to Find the Latest InformationDocumentation is updated periodically. For the latest information about Agilent ESA Spectrum Analyzers, includi
30 Chapter 1 Calibrating Before You Start1.8 GHz Low Pass Filter (for E4404/5/7/8B)Cutoff Frequency: 1.8 GHzRejection at >3 GHz: >45 dB 0955-
300 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BFigure 2-53 Displayed Average Noise Level Tes
Chapter 2 301Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BGenerator) installed and the minimum RBW is 100
302 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BRef Lvl Offset (100 Hz/10 Hz RBW) ____________
Chapter 2 303Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411Bperform those procedures listed in Subset C in
304 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BTable 2-81 DANL Measurement Sequence, E4401B
Chapter 2 305Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BMeasure DANL 10 MHz 500 MHz 1 kHz Off 3)Measure
306 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BMeasuring Displayed Average Noise Level (DANL)
Chapter 2 307Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BHz) if test RBW = 100 Hz or 10 Hz)BW/Avg, Res B
308 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BAMPLITUDE, More, Ref Lvl Offst, (enter Ref Lvl
Chapter 2 309Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403B36. Displayed Average Noise Level: Agilent E440
Chapter 1 31 CalibratingBefore You StartTable 1-11 Recommended Adapters Critical Specifications for Adapter Substitution Recommended HP/Agilent Mo
310 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BFigure 2-54 Displayed Average Noise Level Tes
Chapter 2 311Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403Bcontinue with step 11.6. If the analyzer has a
312 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BFigure 2-54.17.Disconnect the BNC cable and ad
Chapter 2 313Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BSubset D in Table 2-85.8. After performing all
314 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403Bcontinue with “Remove Reference Level Offset”.
Chapter 2 315Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BVideo BW, 1 Hz.8. On the analyzer, press Single
316 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B37. Displayed Average Noise Level: Agilent E44
Chapter 2 317Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BFigure 2-55 Displayed Average Noise Level Test
318 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B6. If the analyzer has a minimum RBW of 1 kHz
Chapter 2 319Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B17.If the analyzer has Option 1DN, 50 Ω trackin
32 Chapter 1 Calibrating Before You StartTable 1-12 Recommended Cables Critical Specifications for Cable SubstitutionRecommended HP/Agilent ModelU
320 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BMeasure DANL 1 GHz 2 GHz 1 kHz Off 2)Measure D
Chapter 2 321Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BMeasuring Displayed Average Noise Level (DANL)U
322 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BRBW = 1 kHz)AMPLITUDE, More, Ref Lvl Offst, (e
Chapter 2 323Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B38. Displayed Average Noise Level: Agilent E440
324 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BFigure 2-56 Displayed Average Noise Level Tes
Chapter 2 325Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B6. If the analyzer does not have Option 1DN (3.
326 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B17.On the analyzer, press Input, Amptd Ref Out
Chapter 2 327Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BSubset D in Table 2-89.Table 2-89 DANL Measure
328 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B8. After performing all applicable DANL measur
Chapter 2 329Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BRBW = 1 kHz)AMPLITUDE, More, Ref Lvl Offst, (en
332 Performance Verification Tests These tests verify the electrical performance of the analyzer. Allow the analyzer to warm up in accordance with the
330 Chapter 2 Performance Verification Tests39. Residual Responses39. Residual ResponsesThe analyzer input is terminated and the analyzer is swept fro
Chapter 2 331Performance Verification Tests39. Residual ResponsesFigure 2-57 Residual Response Test Setup 2. Press Preset on the analyzer. Press the
332 Chapter 2 Performance Verification Tests39. Residual Responsesbeen checked for the model and frequency ranges below.6. Press Single and wait for a
Chapter 2 333Performance Verification Tests40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B,and E4407B (Option AYX)40.
334 Chapter 2 Performance Verification Tests40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option AYX)In
Chapter 2 335Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ
336 Chapter 2 Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and E4411B (Option 1DN or 1
Chapter 2 337Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ
338 Chapter 2 Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and E4411B (Option 1DN or 1
Chapter 2 339Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ
34 Chapter 2 Performance Verification TestsTests included in this section:Tests included in this section:1. 10 MHz Reference Output Accuracy2. 10 MHz
340 Chapter 2 Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4
Chapter 2 341Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B,E4404B, E4405B, E440
342 Chapter 2 Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4
Chapter 2 343Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)43. Tracking Generator
344 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)Additional Equipment f
Chapter 2 345Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)relative to the power l
346 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)5. Disconnect the powe
Chapter 2 347Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)Tracking Generator Leve
348 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)4. Press Single.5. Rec
Chapter 2 349Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)performance verificatio
Chapter 2 35Performance Verification TestsTests included in this section:23. Frequency Response, Agilent E4402B and E4403B24. Frequency Response, Agil
350 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option
Chapter 2 351Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D
352 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option
Chapter 2 353Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D
354 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option
Chapter 2 355Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D
356 Chapter 2 Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)45. Tracki
Chapter 2 357Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Figure 2-65
358 Chapter 2 Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)Source, Am
Chapter 2 359Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)f. Press MKR
36 Chapter 2 Performance Verification TestsTests included in this section:46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4
360 Chapter 2 Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E440
Chapter 2 361Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4408B
362 Chapter 2 Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E440
Chapter 2 363Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4408B
364 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)47. Tra
Chapter 2 365Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Figure 2-
366 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)FREQUEN
Chapter 2 367Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)75 Ω outp
368 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)analyze
Chapter 2 369Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Determini
Chapter 2 37Performance Verification TestsTests included in this section:CalibrationTo perform calibration:1. Run all performance verification tests l
370 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 371Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
372 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 373Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
374 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and
Chapter 2 375Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4
376 Chapter 2 Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B(Option 1
Chapter 2 377Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B(Option 1DN
378 Chapter 2 Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B(Option 1
Chapter 2 379Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option
38 Chapter 2 Performance Verification TestsTests included in this section:Table 2-1 Calibration RequirementsTest #Performance Verification Tests Cali
380 Chapter 2 Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Opti
Chapter 2 381Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option
382 Chapter 2 Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Opti
Chapter 2 383Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option
384 Chapter 2 Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)5
Chapter 2 385Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, andE4407B (Option 1D6)Fig
386 Chapter 2 Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)P
Chapter 2 387Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)52. First LO OUTPUT Power Accuracy (Option AYZ only)Thi
388 Chapter 2 Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)Procedure1. Zero and calibrate the power sensor and po
Chapter 2 389Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)Post-Test Instrument Restoration12.Disconnect the power
Chapter 2 39Performance Verification TestsTests included in this section:24. Frequency Response Frequency Response25. Frequency Response (Preamp On) F
390 Chapter 2 Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)53. IF INPUT Accuracy (Option AYZ only)This test only applies to an
Chapter 2 391Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)Figure 2-73 IF INPUT Accuracy Test SetupProcedureDetermining the IF
392 Chapter 2 Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)6. On the analyzer, press Single and wait for the sweep to finish.7
Chapter 2 393Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)18.Subtract the Input Power (step 14) from the Measured Power (step
394 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)54. Comms Frequency Response (Option BAC or BAH)This test
Chapter 2 395Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Cable, Type-N, 183-cm (72-in) Cable, APC 3.5Termination, 50
396 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)is being measured. 6. Set the source frequency to 50 MHz a
Chapter 2 397Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)810 MHz820 MHz830 MHz840 MHz850 MHz860 MHz880 MHz890 MHz900
398 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)1790 MHz1800 MHz1810 MHz1820 MHz 1830 MHz1840 MHz1850 MHz1
Chapter 2 399Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Part 2: Measuring Frequency Response, 800 MHz to 1000 MHz,
40 Chapter 2 Performance Verification TestsTests included in this section:Manual use of Agilent 8494G and Agilent 8496G Attenuators with the Agilent 1
400 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)splitter. Connect the power splitter to the analyzer 50 Ω
Chapter 2 401Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)17.Perform a peak search on the analyzer. Manually press Se
402 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Table 2-110 Frequency Response WorksheetTest:_______ Atte
Chapter 2 403Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)1740 MHz1750 MHz1760 MHz1770 MHz1780 MHz1790 MHz1800 MHz181
404 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Part 3: Test Results 1. Repeat the following steps using t
Chapter 2 405Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)b. Subtract the Minimum Response for the 1700 MHz to 2000 M
406 Chapter 2 Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)58. GSM Phase and Frequency Error (Options BAH and
Chapter 2 407Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)1. Set up the equipment as shown in Figure 2-76.2. P
408 Chapter 2 Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)14. Repeat step 7 through step 13 for each of the f
Chapter 2 409Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)59. Comms Absolute Power Accuracy (Options BAC or BAH
Chapter 2 41Performance Verification TestsTests included in this section:Table 2-2 Agilent 11713A Settings for Agilent 8494G and Agilent 8496G1 dB St
410 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Additional Equipment for Option BABAdapter, Type N (
Chapter 2 411Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)1. Press Preset on the analyzer. Press the Factory Pr
412 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)0.02 dB, then 1 dB Actual0dB is 0.02 dB. In this cas
Chapter 2 413Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Figure 2-78 Amplitude Accuracy Test Setup 12.Set the
414 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)14.Set the 10 dB and 1 dB step attenuators to 0 dB.1
Chapter 2 415Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Table 2-120 GSM Transmit Power (Option BAH), Preamp
416 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)19.Table 2-117 and Table 2-121 have two Amplitude Ac
Chapter 2 417Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)numbers in parenthesis.25.Continue with Part 2.Table
418 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Table 2-125 GSM Transmit Power Accuracy (Cellular B
Chapter 2 419Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Part 2: cdmaOne Receive Channel Power (Preamp Off)Tab
42 Chapter 2 Performance Verification Tests1. 10 MHz Reference Output Accuracy1. 10 MHz Reference Output AccuracyThe setability is measured by changin
420 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)NOTE The external attenuators and cables are now par
Chapter 2 421Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)h. Calculate the amplitude accuracy as follows:7. Tab
422 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)to 1 GHz Cellular Band, and the 1.7 GHz to 2 GHz PCS
Chapter 2 423Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Part 3: cdmaOne Receive Channel Power (Preamplifier O
424 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)4. Calculate the actual total attenuation by adding
Chapter 2 425Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)f. Press Peak Search (or Search). The marker can stil
426 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)−90 0 0.24 39)−100 0 0.24 40)Table 2-134 cdmaOne Re
4273 Performance Verification Test Records
428 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordAgilent E4401B Performance Verification Test Re
Chapter 3 429Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record50 Ω Termination ___________ ___________ ______
Chapter 2 43Performance Verification Tests1. 10 MHz Reference Output Accuracy10 s.d. Press Enter.e. On Channel 1, press 50 Ω/1 MΩ until the LED is lit
430 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 1490 MHz 10 MHz 1489.91 MHz (2)________
Chapter 3 431Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordOffset from 500 MHz signal 30 kHz to 230 kHz (1
432 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–12 –0.40 dB (3)________ 0.40 dB ±0.075 dB–16
Chapter 3 433Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–20 –0.4 dB (26)________ 0.4 dB ±0.078 dB–24 –0.
434 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–24 –0.54 dB (48)________ 0.54 dB ±0.074 dB–28
Chapter 3 435Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–12 –0.4 dB (70)________ 0.4 dB ±0.075 dB–16
436 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–20 –2.0% (97)________ 2.0% ±0.063%Linear Fidel
Chapter 3 437Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–56 –1.14 dB (116)________ 1.14 dB ±0.078 dB–60
438 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–5 43.75 –0.3 dB (2)________ 0.3 dB ±0.100 dB–3
Chapter 3 439Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–75 –26.25 –0.7 dB (21)________ 0.7 dB ±0.099 dB
44 Chapter 2 Performance Verification Tests1. 10 MHz Reference Output Accuracythe maximum frequency change.16.Divide the maximum frequency change by t
440 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 200 Hz (Option 1DR only) –0.3 dB (12)________
Chapter 3 441Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record –20 dBm input –0.54 dB (7)________ 0.54 dB
442 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record20 to 30° C50 Ω Input (Non-Option 1DP)(9 kHz to
Chapter 3 443Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record25. Frequency Response (Preamp On)(Option 1DS on
444 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record Peak-to-Peak Response (3)________ 4.0 dB ±0.
Chapter 3 445Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 53 MHz (1)________ 1.0 dB ±0.123 dB 50.004 MHz
446 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 50 Ω, 10 Hz RBW, Preamp On 400 kHz (16)_
Chapter 3 447Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 1 MHz to 10 MHz (33)________ –99 dBmV ±1
448 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordPower Sweep Accuracy (4)________ 1.8 dB ±0.008
Chapter 3 449Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record2nd Harmonic, ≥20 MHz(2)________ –25 dBc±2.6 dB3
Chapter 2 45Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B,E4404B, E4405B, and E44
450 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordAgilent E4402B Performance Verification Test Re
Chapter 3 451Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record10 dB Step Attenuator ___________ ___________ __
452 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Center Freq Span 1500 MHz 10 MHz 14
Chapter 3 453Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record (Serial Prefix ≥ US4510, SG4510 or MY4510, Non
454 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record 1 kHz Res BW,(Non-Option 1D5) (1)_________ 15
Chapter 3 455Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–24 –0.50 dB (6)________ 0.50 dB ±0.074 dB–28
456 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–32 –0.4 dB (29)________ 0.4 dB ±0.077 dB–36
Chapter 3 457Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–40 –0.70 dB (52)________ 0.70 dB ±0.081 dB–44
458 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–28 –0.4 dB (74)________ 0.4 dB ±0.073 dB–32 –0
Chapter 3 459Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–8 –2.0% (99)________ 2.0% ±0.064%–12 –2.0% (100
46 Chapter 2 Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E
460 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record13. Input Attenuation Switching UncertaintyInpu
Chapter 3 461Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–70 dBm –0.5 dB (7)________ 0.5 dB ±0.099 dB–80
462 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–10 dBm –0.3 dB (25)________ 0.3 dB ±0.091 dB 0
Chapter 3 463Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record 3 Hz (Option 1DR and 1D5 only) Firmware Revisio
464 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−40 dBm Reference Level –40 dBm input –0.54
Chapter 3 465Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Minimum Response –0.46 dB (2)________ ±0.19 d
466 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordOption UKBac CoupledBand 0(100 kHz to 3.0 GHz)
Chapter 3 467Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Peak-to-Peak Response (3)________ 1.52 dB ±0.
468 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordOption UKBac CoupledBand 0(100 kHz to 3.0 GHz)
Chapter 3 469Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Minimum Response –1.5 dB (2)________ ±0.28 dB
Chapter 2 47Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B,E4404B, E4405B, and E44
470 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordBand 0A(800 MHz to 1.0 GHz)(serial number US394
Chapter 3 471Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record33. Gain CompressionTest Frequency 53 MHz (1)__
472 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record1kHz RBW, Preamp On,20 to 30° C 10 MHz
Chapter 3 473Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–8 dB –0.5 dB (7)________ 0.5 dB ±0.008 dB–9 dB
474 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Positive Level Flatness, >1.5 GHz(7)_____
Chapter 3 475Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record50. Gate Delay Accuracy and Gate Length Accuracy
476 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−45 dBm –0.70 dB (4)________ 0.70 dB ±0.152 dB−
Chapter 3 477Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−30 dBm–0.75 dBa(21)________0.75 dBb±0.156 dB−40
478 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−40 dBm –1.77 dB (35)________ 1.77 dB ±0.161 dB
Chapter 3 479Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordAgilent E4403B Performance Verification Test Rec
48 Chapter 2 Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B3. Freq
480 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordMicrowave Spectrum Analyzer (Option 1DN only) _
Chapter 3 481Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record6. Frequency Span Accuracy Span Start Fr
482 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record 20 ms –1.0% (2)________ 1.0% ±0.16% 100 ms
Chapter 3 483Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–76 –1.06 dB (19)________ 1.06 dB ±0.083 dB–80
484 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≤300 Hz(Option 1
Chapter 3 485Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–92 –1.22 dB (65)________ 1.22 dB ±0.100 dB–96
486 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–80 –0.4 dB (87)________ 0.4 dB ±0.089 dBLinear
Chapter 3 487Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–28 –0.72 dB (109)________ 0.72 dB ±0.073 dB–32
488 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB65
Chapter 3 489Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordResolution Bandwidth 3 kHz –0.3 dB (1)________ 0
Chapter 2 49Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B,E4403B, and E4411B“Part 2:
490 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record –40 dBm input –0.6 dB (5)________ 0.6 dB ±
Chapter 3 491Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record23. Frequency Response Note: Enter data in the
492 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record36. Displayed Average Noise Level1 kHz RBW
Chapter 3 493Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordIncremental Vernier Accuracy–2 dB –0.2 dB (10)__
494 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record46. Tracking Generator Harmonic Spurious Output
Chapter 3 495Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordAgilent E4404B Performance Verification Test Rec
496 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordPower Splitter ___________ ___________ ________
Chapter 3 497Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1
498 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record7. Noise Sidebands Offset from 1 GHz signal (
Chapter 3 499Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Option 120 Offset from 1 GHz signal 100
Contents 51. CalibratingCalibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
50 Chapter 2 Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411Bdisplay
500 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 1 ms (Option AYX or B7D only)–1.0% (6)______
Chapter 3 501Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–72 –0.80 dB (18)________ 0.80 dB ±0.084 dB–76
502 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–80 –0.4 dB (41)________ 0.4 dB ±0.089 dBCumula
Chapter 3 503Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–88 –1.18 dB (64)________ 1.18 dB ±0.085 dB–92
504 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–76 –0.4 dB (86)________ 0.4 dB ±0.083
Chapter 3 505Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–24 –0.66 dB (108)________ 0.66 dB ±0.074 dB–28
506 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record55 dB –0.65 dB (11)________ 0.65 dB ±0.099 dB60
Chapter 3 507Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordLog, Res BW ≤300 Hz(Option 1DR only)Reference Le
508 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 10 kHz –0.3 dB (3)________ 0.3 dB ±0.048 dB 30
Chapter 3 509Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record0 dBm Reference Level 0 dBm input –0.54 dB (
Chapter 2 51Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,E4407B, and E4408B4. Freque
510 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record10 kHz 8.5 kHz (7)________ 11.5 kHz ±220 Hz3 kH
Chapter 3 511Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Peak-to-Peak Response (6)________ 2.6 dB ±0.3
512 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Maximum Response (19)________ 1.5 dB ±0.39 d
Chapter 3 513Performance Verification Test RecordsAgilent E4404B Performance Verification Test Recordac CoupledBand 0(100 kHz to 3.0 GHz) Maximum R
514 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Peak-to-Peak Response (21)________ 3.0 dB ±0
Chapter 3 515Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record0 to 55° CBand 0(1 MHz to 3.0 GHz) Maximum Res
516 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 2.0 GHz 5600.0 MHz (5)________ –80 dBc ±.8
Chapter 3 517Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 10 MHz to 1 GHz (1)________ –116 dBm ±
518 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record10 Hz RBW, Preamp On,20 to 30° C 10 MH
Chapter 3 519Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–2 dB –0.2 dB (10)________ 0.2 dB ±0.008 dB–3 dB
52 Chapter 2 Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408BProcedu
520 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record46. Tracking Generator Harmonic Spurious Output
Chapter 3 521Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record65 ms Gate Length 64.993 ms (4)________ 65.007 m
522 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−25 dBm –0.74 dB (9)________ 0.74 dB ±0.161 dB−
Chapter 3 523Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−40 dBm –1.46 dB (25)________ 1.46 dB ±0.161 dB−
524 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−70 dBm –1.88 dB (43)________ 1.88 dB ±0.161 dB
Chapter 3 525Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordAgilent E4405B Performance Verification Test Rec
526 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordPower Splitter ___________ ___________ ________
Chapter 3 527Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1
528 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Center Freq Span 1500 MHz 20 MHz 149
Chapter 3 529Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 20 kHz (2)________ –100 dBc/Hz ±2.44 dB
Chapter 2 53Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,E4407B, and E4408BPart 2: M
530 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 10 MHz (4)________ –137 dBc/Hz ±2.44 dB9
Chapter 3 531Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≥1 kHzdB from Ref
532 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≥1 kHzdB from R
Chapter 3 533Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB–8 –0
534 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≤300 Hz(Option
Chapter 3 535Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–4 –2.0% (93)________ 2.0% ±0.064%–8 –2.0% (94)_
536 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–40 –0.90 dB (112)________ 0.90 dB ±0.081 dB–44
Chapter 3 537Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record15. Reference Level AccuracyLog, Res BW ≥1 kHzRe
538 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 0 dBm –0.3 dB (18)________ 0.3 dB ±0.100 dB–30
Chapter 3 539Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB 30
54 Chapter 2 Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408Bverific
540 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record –10 dBm input –0.54 dB (2)________ 0.54 dB
Chapter 3 541Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz120 kHz
542 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Recorddc CoupledBand 2(6.7 GHz to 13.2 GHz) Maximum
Chapter 3 543Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Minimum Response –0.5 dB (17)________ ±0.11 d
544 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordOption UKBdc CoupledBand 0(100 Hz to 3.0 GHz)
Chapter 3 545Performance Verification Test RecordsAgilent E4405B Performance Verification Test Recordac CoupledBand 0A(800 MHz to 1.0 GHz)(serial num
546 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Peak-to-Peak Response (24)________ 4.0 dB ±0
Chapter 3 547Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record0 to 55° CBand 0(1 MHz to 3.0 GHz) Maximum Res
548 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 2.0 GHz 5600.0 MHz (5)________ –80 dBc ±.8
Chapter 3 549Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 1403 MHz (3)________ 1.0 dB ±0.123 dB 2503 MHz
Chapter 2 55Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411B5. Frequency Span Accuracy: Agilent E4401B and E4411BFo
550 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record10 Hz RBW, Preamp On,0 to 55° C 10 MHz
Chapter 3 551Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB–3 dB
552 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Negative Level Flatness, 1 MHz to 10 MHz–3.0
Chapter 3 553Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 9 kHz to 2.9 GHz (1)________ –16 dBm ±2.58 dB 2
554 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−5 dBm –0.82 dB (2)________ 0.82 dB ±0.161 dB−2
Chapter 3 555Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−20 dBm –0.77 dB (20)________ 0.77 dB ±0.161 dB−
556 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−40 dBm –1.88 dB (35)________ 1.88 dB ±0.161 dB
Chapter 3 557Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordAgilent E4407B Performance Verification Test Rec
558 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordPower Splitter ___________ ___________ ________
Chapter 3 559Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1
56 Chapter 2 Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411BFigure 2-5 Frequency Span Readout Accuracy Test Setu
560 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 1500 MHz 1 MHz 1499.999998 MHz(17)_____
Chapter 3 561Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 9000 MHz 1 MHz 8999.999999 MHz(21)______
562 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record (Serial Prefix ≥ US4510, SG4510 or MY4510, Op
Chapter 3 563Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record9. System-Related Sidebands Offset from 500 MHz
564 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≥1 kHzdB from Re
Chapter 3 565Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≥1 kHzdB from Re
566 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB–8 –0
Chapter 3 567Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≤300 Hz(Option 1
568 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–4 –2.0% (93)________ 2.0% ±0.064%–8 –2.0% (94)
Chapter 3 569Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–40 –0.90 dB (112)________ 0.90 dB ±0.081 dB–44
Chapter 2 57Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411BCW, 100 MHzPOWER LEVEL, –5 dBm3. Set the synthesized si
570 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record15. Reference Level AccuracyLog, Res BW ≥1 kHzR
Chapter 3 571Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 0 dBm –0.3 dB (18)________ 0.3 dB ±0.100 dB–30
572 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB 3
Chapter 3 573Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record –10 dBm input –0.54 dB (2)________ 0.54 dB
574 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz120 kHz
Chapter 3 575Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Peak-to-Peak Response (6)________ 0.92 dB ±0.
576 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Minimum Response –2.0 dB (17)________ ±0.59
Chapter 3 577Performance Verification Test RecordsAgilent E4407B Performance Verification Test Recordac CoupledBand 1(3.0 GHz to 6.7 GHz) Maximum R
578 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBdc CoupledBand 0(100 Hz to 3.0 GHz)
Chapter 3 579Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Maximum Response (10)________ 2.5 dB ±0.39 dB
58 Chapter 2 Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411B10 110 20 100 2)10010 10.1 10.01 10.09 3)0.1800 900 81
580 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0A(800 MHz to 1.0 GHz)
Chapter 3 581Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Minimum Response –3.0 dB (32)________ ±0.40 d
582 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordNon-Option UKBdc CoupledBand 0B(1.7 GHz to 2.0
Chapter 3 583Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0B(1.7 GHz to 2.0 GHz)(
584 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordNon-Option UKBdc CoupledBand 0B(1.7 GHz to 2.0
Chapter 3 585Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0B(1.7 GHz to 2.0 GHz)(
586 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 15.0 GHz 18830.35 MHz (17)________ –65 dBc ±
Chapter 3 587Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 14003 MHz (7)________ 1.0 dB ±0.123 dB38. Disp
588 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 10 MHz to 1 GHz (18)________ –150 dBm
Chapter 3 589Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–5 dB –0.5 dB (4)________ 0.5 dB ±0.008 dB–6 dB
Chapter 2 59Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B6. Frequency Span Accu
590 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Positive Level Flatness, >10 MHz to 1.5 G
Chapter 3 591Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record50. Gate Delay and Gate Length Accuracy (Option
592 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record6.9 GHz 13.0 dBm (13)________ 17.5 dBm ±0.16 dB
Chapter 3 593Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record58. GSM Phase and Frequency Error (Option BAH an
594 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordGSM Transmit Power Accuracy (Option BAH only)GS
Chapter 3 595Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record−85 dBm –1.56 dB (29)________ 1.56 dB ±0.152 dBP
596 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordAgilent E4408B Performance Verification Test Re
Chapter 3 597Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record1 dB Step Attenuator ___________ ___________ ___
598 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 16000 MHz 20 MHz 15999.83 MHz(10)______
Chapter 3 599Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 16000 MHz 1 MHz 20999.999995 MHz(23)____
6 ContentsE4405B, E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13619. Overall Absol
60 Chapter 2 Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BFigure 2-6 Frequenc
600 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record Span Start Freq 3000 MHz 0 Hz 2370
Chapter 3 601Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 1 s –1.0% (4)________ ±1.0% ±0.16% 10 s –1.
602 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–84 –1.14 dB (21)________ 1.14 dB ±0.086 dBIn
Chapter 3 603Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≤300 Hz(Option 1D
604 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–92 –1.22 dB (65)________ 1.22 dB ±0.100 dB–96
Chapter 3 605Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–80 –0.4 dB (87)________ 0.4 dB ±0.089 dBLinear
606 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–28 –0.72 dB (109)________ 0.72 dB ±0.073 dB–32
Chapter 3 607Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB65
608 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordResolution Bandwidth 3 kHz –0.3 dB (1)________
Chapter 3 609Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record –40 dBm input –0.6 dB (5)________ 0.6 dB ±
Chapter 2 61Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BPOWER LEVEL, –5 dBm3.
610 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record24. Frequency Response Note: Enter data in the
Chapter 3 611Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record0 to 55° CBand 0(9 kHz to 3.0 GHz Maximum Resp
612 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 2.0 GHz 2042.8 MHz (1)________ –65 dBc ±.8
Chapter 3 613Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 8 GHz TOI 5 dBm (4)________ ±0.93 dB 300 MHz
614 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 2 GHz to 3 GHz (13)________ –120 dBm
Chapter 3 615Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–6 dB –0.2 dB (13)________ 0.2 dB ±0.008 dB–7 dB
616 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record3rd Harmonic, <20 kHz(3)________ –15 dBc±2.6
Chapter 3 617Performance Verification Test RecordsAgilent E4411B Performance Verification Test RecordAgilent E4411B Performance Verification Test Rec
618 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record50Ω Termination ___________ ___________ ______
Chapter 3 619Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record 1490 MHz 1 MHz 1489.999998 MHz(5)_______
62 Chapter 2 Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BTable 2-9 Frequency
620 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Sweep Time 5 ms –1.0% (1)________ 1.0%±0.
Chapter 3 621Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–68 –0.98 dB (17)________ 0.98 dB ±0.080 dB–72
622 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–76 –0.4 dB (40)________ 0.4 dB ±0.083
Chapter 3 623Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–84 –1.14 dB (63)________ 1.14 dB ±0.086 dB–88
624 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–72 –0.4 dB (85)________ 0.4 dB ±0.084 dB–76
Chapter 3 625Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–20 –0.60 dB (107)________ 0.60 dB ±0.078 dB–24
626 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record50 dB –0.60 dB (10)________ 0.60 dB ±0.095 dB5
Chapter 3 627Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record16. Resolution Bandwidth Switching UncertaintyRe
628 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record –30 dBm input –0.6 dB (4)________ 0.6 dB
Chapter 3 629Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record9 kHz 7.2 kHz (11)________ 10.8 kHz ±163 Hz22. F
Chapter 2 63Performance Verification Tests7. Noise Sidebands7. Noise SidebandsA 1 GHz CW signal is applied to the input of the analyzer. The marker fu
630 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Minimum Response –1.0 dB (2)________ ±0.11 d
Chapter 3 631Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record 50 Ω, 100 Hz RBW 400 kHz (11)________ –12
632 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test RecordNegative Vernier Accuracy–0.75 dB (3)________ ±
Chapter 3 633Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Positive Level Flatness, >10 MHz(5)_______
634 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record
6354 If You Have a ProblemThis chapter includes information on how to check for a problem with your Agilent ESA Series Spectrum Analyzer, and how to r
636 Chapter 4 If You Have a ProblemWhat You’ll Find in This ChapterWhat You’ll Find in This Chapter This chapter includes information on how to check
Chapter 4 637If You Have a ProblemBefore You Call Agilent TechnologiesBefore You Call Agilent Technologies Check the Basics A problem can often be res
638 Chapter 4 If You Have a ProblemBefore You Call Agilent Technologieschapter.❏ If the equipment to perform the performance verification tests is not
Chapter 4 639If You Have a ProblemBefore You Call Agilent TechnologiesTable 4-1 Agilent Technologies Sales and Service OfficesUNITED STATESInstrument
64 Chapter 2 Performance Verification Tests7. Noise SidebandsProcedure This performance test consists of five parts: Part 1: Noise Sideband Suppressio
640 Chapter 4 If You Have a ProblemHow to Return Your Analyzer for ServiceHow to Return Your Analyzer for Service Service Tag If you are returning you
Chapter 4 641If You Have a ProblemHow to Return Your Analyzer for ServiceFigure 4-1
642 Chapter 4 If You Have a ProblemHow to Return Your Analyzer for ServiceOther Packaging CAUTION Analyzer damage can result from using packaging mate
Chapter 2 65Performance Verification Tests7. Noise SidebandsDet/Demod, Detector, SampleSweep, Sweep Time, 5 secSingle Wait for the completion of a swe
66 Chapter 2 Performance Verification Tests7. Noise Sidebands4. Press the following analyzer keys to measure the carrier amplitude:Peak Search (or Sea
Chapter 2 67Performance Verification Tests7. Noise SidebandsSweep, Sweep Time, 5 secSingleWait for the completion of a sweep, then press Peak Search (
68 Chapter 2 Performance Verification Tests7. Noise SidebandsFREQUENCY, ↓, ↓Single Record the marker amplitude noise reading in Table 2-10 as the Lowe
Chapter 2 69Performance Verification Tests7. Noise Sidebandsverification test record in locations 1, 2, 3 and 4 respectively.6. This completes the Noi
Contents 742. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN) . .
70 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B8. Noise Sidebands - Wide Offse
Chapter 2 71Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407Bnoise sideband suppression. Part
72 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407BNOTE A spur may exist at 100 kH
Chapter 2 73Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407BPart 3: Noise Sideband Suppressi
74 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407Breading in Table 2-11 as the Lo
Chapter 2 75Performance Verification Tests9. System-Related Sidebands9. System-Related SidebandsA 500 MHz CW signal is applied to the input of the ana
76 Chapter 2 Performance Verification Tests9. System-Related Sidebandsa. Set the synthesized signal generator controls as follows: FREQUENCY, 500 MHz
Chapter 2 77Performance Verification Tests9. System-Related SidebandsSingle. Wait for the completion of a new sweep, then press Peak Search (or Search
78 Chapter 2 Performance Verification Tests10. Residual FM10. Residual FMThis test measures the inherent short-term instability of the analyzer LO sys
Chapter 2 79Performance Verification Tests10. Residual FMCAUTION Use only 75 Ω cables, connectors, or adapters on instruments with 75 Ω connectors, or
8 ContentsAgilent E4403B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .479Agilent E4404B Performance Ver
80 Chapter 2 Performance Verification Tests10. Residual FMdelta (∆ Mkr1) amplitude reads −8dB ±0.3 dB. 7. Press Delta, then rotate the knob counterclo
Chapter 2 81Performance Verification Tests10. Residual FMstep 8 by the Deviation recorded in step 12. Record this value as Test Record entry 1 (Residu
82 Chapter 2 Performance Verification Tests10. Residual FMreference level. If it is not, press Sweep, Sweep (Cont), FREQUENCY, and use the knob to pla
Chapter 2 83Performance Verification Tests11. Sweep Time Accuracy11. Sweep Time AccuracyThis test uses a function generator to amplitude-modulate a 50
84 Chapter 2 Performance Verification Tests11. Sweep Time Accuracyconnectors, or the connectors will be damaged.Procedure 1. Set the synthesized signa
Chapter 2 85Performance Verification Tests11. Sweep Time Accuracyaccuracy as follows:NOTE The sweep time accuracy is defined as a percentage of the in
86 Chapter 2 Performance Verification Tests12. Display Scale Fidelity12. Display Scale FidelityA 50 MHz CW signal is applied to the input of the analy
Chapter 2 87Performance Verification Tests12. Display Scale FidelityFigure 2-12 Scale Fidelity Test Setup CAUTION Use only 75 Ω cables, connectors, a
88 Chapter 2 Performance Verification Tests12. Display Scale FidelityIf the calibration data does not indicate an actual attenuation value for the 0 d
Chapter 2 89Performance Verification Tests12. Display Scale Fidelity6. Press Peak Search (or Search) on the analyzer.7. Adjust the synthesized signal
91Calibrating
90 Chapter 2 Performance Verification Tests12. Display Scale FidelityCalculate Incremental Log Fidelity1. Calculate the Incremental Log Fidelity Error
Chapter 2 91Performance Verification Tests12. Display Scale FidelityExample Calculation for ILFE at –20 dB from REFLVL setting:2. Record the result in
92 Chapter 2 Performance Verification Tests12. Display Scale FidelityBW/Avg, Average (On)5, Enter3. Press Peak Search (or Search) on the analyzer and
Chapter 2 93Performance Verification Tests12. Display Scale FidelityCalculate incremental log fidelity:Calculate the Incremental Log Fidelity Error (I
94 Chapter 2 Performance Verification Tests12. Display Scale FidelitySPAN, 10 kHzBW/Avg, 3 kHzAMPLITUDE, Scale Type (Lin)Input/Output (or Input), Inpu
Chapter 2 95Performance Verification Tests12. Display Scale FidelityTable 2-15.11.Calculate the Linear Fidelity Error (LFE) as a percentage of referen
96 Chapter 2 Performance Verification Tests12. Display Scale FidelitySetup for linear scale measurement:1. Set the following parameters on the analyze
Chapter 2 97Performance Verification Tests12. Display Scale Fidelity10.If the analyzer has a 75W Input press:Input/Output (or Input)Input Z Corr (75)Z
98 Chapter 2 Performance Verification Tests12. Display Scale Fidelity16. Press SPAN, Zero Span on the analyzer.17.Adjust the synthesized signal genera
Chapter 2 99Performance Verification Tests12. Display Scale FidelityPost-Test Instrument Restoration 25. Remove the RF cable from the analyzer input