Chapter 2 37
Performance Verification Tests
Tests included in this section:
Calibration
To perform calibration:
1. Run all performance verification tests listed in Column 1 of Table 2-1
that are applicable to the analyzer being calibrated and the installed
options. Refer to the Performance Verification test tables in
Chapter 1 for a list of the tests that are applicable to the analyzer
being calibrated.
2. If any of the performance verification tests fail, perform the
appropriate calibration adjustments listed in Column 2 of Table 2-1
which corresponds to the failure.
3. Repeat all of the performance verification tests listed in Column 1 of
Table 2-1 if any calibration adjustments were made in step 2. This
will confirm that there is no interaction between adjustments which
could negatively impact analyzer performance.
Table 2-1 lists the performance verification tests and adjustments
needed for calibration. The performance tests are located in the
following pages of this chapter. Adjustment information is located in the
service guide.