Chapter 3 543
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
Minimum Response –0.5 dB (17)________ ±0.11 dB
Peak-to-Peak Response (18)________ 1.0 dB
±0.11 dB
ac Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (19)________ 1.5 dB
±0.39 dB
Minimum Response –1.5 dB (20)________
±0.39 dB
Peak-to-Peak Response (21)________ 2.6 dB
±0.39 dB
ac Coupled
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (22)________ 2.0 dB
±0.40 dB
Minimum Response –2.0 dB (23)________
±0.40 dB
Peak-to-Peak Response (24)________ 3.6 dB
±0.40 dB
0 to 55
° C
Non-Option UKB
dc Coupled
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.76 dB
±0.19 dB
Minimum Response –0.76 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 1.52 dB
±0.19 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty