Agilent Technologies E1401-90493 Manuale Utente Pagina 264

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 642
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 263
264 Chapter 2
Performance Verification Tests
31. Spurious Responses: Agilent E4402B and E4403B
Adapter, Type-N (m) to BNC (f)
Adapter, Type-N (m) to Type-N (m)
Adapter, SMA (f) to BNC (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Procedure
This performance test consists of two parts:
Part 1: Third Order Intermodulation Distortion
Part 2: Second Harmonic Distortion
Perform Part 1 before Part 2.
Part 1: Third Order Intermodulation Distortion
1. Zero and calibrate the power meter and RF power sensor in log mode
(power reads out in dBm), as described in the power meter operation
manual.
2. Connect the equipment as shown in Figure 2-47 with the output of
the directional bridge connected to the power sensor.
Figure 2-47 Third Order Intermodulation Distortion Test Setup
Vedere la pagina 263
1 2 ... 259 260 261 262 263 264 265 266 267 268 269 ... 641 642

Commenti su questo manuale

Nessun commento