Agilent Technologies E1401-90493 Manuale Utente Pagina 582

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 642
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 581
582 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Non-Option UKB
dc Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (7)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (8)________
±0.28 dB
Peak-to-Peak Response (9)________ 3.0 dB
±0.28 dB
Option UKB
ac Coupled
Band 0
(10 MHz to 3.0 GHz)
Maximum Response (10)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (11)________
±0.28 dB
Peak-to-Peak Response (12)________ 3.0 dB
±0.28 dB
Option UKB
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (13)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (14)________ ±0.28 dB
Peak-to-Peak Response (15)________ 3.0 dB
±0.28 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Vedere la pagina 581
1 2 ... 577 578 579 580 581 582 583 584 585 586 587 ... 641 642

Commenti su questo manuale