Agilent Technologies N9340A Manuale Utente Pagina 115

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 172
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 114
SCPI Command Reference 6
N9340A User’s Guide 105
Control Limit Pattern Testing
:CALCulate:LLINe[1]:PATTern[:STATe]
OFF|ON|0|1
:CALCulate:LLINe[1]:PATTern[:STATe]?
This command toggles the usage of limit pattern
between on and off.
If the limit pattern is turned on, limit line is
automatically turned off.
Set Limit Pattern X-axis Units
:CALCulate:LLINe[1]:PATTern:DOMain
FREQ|TIME
:CALCulate:LLINe[1]:PATtern:DOMain?
This command selects how the limit pattern points
are defined: according to the frequency or sweep
time setting of the spectrum analyzer. The query
command returns a string of FREQ or TIME.
*RST state: Off
Key access: [Limit] > {Limit Pattern}
*RST state: Frequency
Range: FREQ|TIME
Key access: [Limit] > {Set pattern} > {X axis unit}
Vedere la pagina 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 171 172

Commenti su questo manuale

Nessun commento