SCPI Command Reference 6
N9340A User’s Guide 123
SENSe Subsystem
Sets the instrument parameters for the input signal
measurements.
[:SENSe]:AVERage Subsection
Average Type
[:SENSe]:AVERage:TYPE
LOGPower|POWer|VOLTage
[:SENSe]:AVERage:TYPE?
This command sets the average type to the
following three types:
Log Pwr
Power
Voltage
For more information about the three types of
average above, please refer to “Avg Type" on page 74
Clear the current Average
[:SENSe]:AVERage:TRACe1|2|3|4|5|6:CLEar
This command restarts the trace averaging
function.
*RST State: Log Pwr
Range: Log Pwr, Pwr, Voltage
Key Access: [BW/SWP] > {Avg Type}
NOTE
Restart the trace at the beginning of a sweep to obtain valid
average data. To do this remotely abort the sweep and initiate
a single sweep.